Instruction manual
Chapter 4 Command Reference
[:SENSe]:FUNCtion Subtree
4-30 Programming Guide
4
[:SENSe]:ROSCillator Subtree
This subtree controls the Reference Oscillator.
COMMAND [:SENSe]:ROSCillator:SOURce . . . <character_program_data>
Sets or queries the current reference timebase to INTernal or EXTernal.
QUERY RESPONSE A sequence of ASCII-encoded bytes: INT or EXT
COMMENTS • *RST: INT
• INTernal indicates the timebase is the internal reference. EXTernal
indicates the signal at the external reference input (located on the rear
panel of the Counter; Reference connector) is the reference timebase.
• The Counter does not switch to EXTernal unless a suitable 1, 2, 5, or
10 MHz signal is present on the “Reference” connector on the back panel.
• Execution of the command (that is, explicitly selecting internal or
external timebase) sets [:SENS]:ROSC:SOUR:AUTO to OFF.
• If this is set to EXT, and no valid external signal is available at the
back-panel Reference connector, the front-panel frequency display may
show an error message.
• If this is set to INT, the 10 MHz signal generated by the internal
reference oscillator is available as an output on the back-panel
Reference connector.
RELATED
FRONT-PANEL
KEYS
Shift + Menu (REF OSC > INT | EXT)