Chapter 4 Command Reference
[:SENSe] Subsystem
4-24 Programming Guide
4
COMMAND [:SENSe]:DATA? . . . [<data handle>]
Queries the current measurement result data of the :SENSe subsystem.
Valid <data handles>:
“[SENSe:][XNONE:]FREQuency [1] | 2”
“[SENSe:][XNONE:]POWer [2]”
QUERY RESPONSE Frequency values are returned in Hz as ASCII bytes in NR1 format.
Power values are returned in dB in NR2 format.
COMMENTS Query only.
Does not initiate any measurement action.
The data handle is optional. If it is omitted, the query returns values
for all functions enabled by the FUNCtion:ON command.
Values are returned in the same order as the functions returned by the
FUNCtion:ON? query.
If this query executes while a measurement is in progress, the prior
measurement result is returned, if the prior result was not invalidated.
RELATED
FRONT-PANEL
KEYS
None
COMMAND [:SENSe]:FILTer:FM:AUTO . . . <Boolean>
Turns the Counter’s ability to automatically compensate for frequency
modulation ON or OFF. (*RST: ON)
COMMENTS When FM:AUTO is ON, the Counter automatically detects FM signals
and modifies its measurement algorithm accordingly.
Using FM:AUTO increases the time required to compute each
measurement but increases the accuracy of FM signal measurements.
RELATED
FRONT-PANEL
KEYS
Shift + Menu (FM > AUTO | OFF)