Instruction manual
Chapter 4 Command Reference
[:SENSe] Subsystem
4-24 Programming Guide
4
COMMAND [:SENSe]:DATA? . . . [<data handle>]
Queries the current measurement result data of the :SENSe subsystem.
Valid <data handles>:
“[SENSe:][XNONE:]FREQuency [1] | 2”
“[SENSe:][XNONE:]POWer [2]”
QUERY RESPONSE • Frequency values are returned in Hz as ASCII bytes in NR1 format.
• Power values are returned in dB in NR2 format.
COMMENTS • Query only.
• Does not initiate any measurement action.
• The data handle is optional. If it is omitted, the query returns values
for all functions enabled by the FUNCtion:ON command.
• Values are returned in the same order as the functions returned by the
FUNCtion:ON? query.
• If this query executes while a measurement is in progress, the prior
measurement result is returned, if the prior result was not invalidated.
RELATED
FRONT-PANEL
KEYS
None
COMMAND [:SENSe]:FILTer:FM:AUTO . . . <Boolean>
Turns the Counter’s ability to automatically compensate for frequency
modulation ON or OFF. (*RST: ON)
COMMENTS • When FM:AUTO is ON, the Counter automatically detects FM signals
and modifies its measurement algorithm accordingly.
• Using FM:AUTO increases the time required to compute each
measurement but increases the accuracy of FM signal measurements.
RELATED
FRONT-PANEL
KEYS
Shift + Menu (FM > AUTO | OFF)