Chapter 4 Command Reference
:MEMory Subsystem
Programming Guide 4-21
4
COMMAND :MEMory:CLEAR[:NAME] . . . <name>
Resets the contents of the named power-correction profile to the default
configuration.
COMMENTS • Valid profile names: CORR1, CORR2, . . . CORR9
• The data currently stored in non-volatile memory for the named
correction profile is discarded.
• The default power-correction profile configuration consists of two data
points, both having loss values of zero. One of these points contains a
frequency value of 1 GHz, and the other contains a frequency value
equal to the highest frequency the Counter model can measure
(53150A: 20 GHz; 53151A: 26.5 GHz; 53152A: 46 GHz). The remaining
eight data points contain loss and frequency values of zero.
• This is an event. There is no query form of this command.
RELATED
FRONT-PANEL
KEYS
Shift + Menu (PWR CORR)
COMMAND :MEMory:NSTates?
Queries the Number of available *SAV/*RCL STates in the instrument.
QUERY RESPONSE • Numeric data transferred as ASCII bytes in <NR1> format.
• The value returned is 9.
• The response value is one greater than the maximum which can be
sent as a parameter to the *SAV and *RCL commands.
COMMENTS Query only.
RELATED
FRONT-PANEL
KEYS
None