Instruction manual
Chapter 4 Command Reference
:MEMory Subsystem
4-20 Programming Guide
4
:MEMory Subsystem
This subsystem manages the instrument’s memory.
COMMAND :MEMory:DATA . . . <name>,<data>
Stores and queries data in the named power-correction profile.
QUERY RESPONSE • Data points are returned in <definite length arbitrary block> format.
• A data point consists of two to ten comma-separated, NRf format
number pairs.
COMMENTS • Valid profile names: CORR1, CORR2, . . . CORR9
• Valid data is stored in non-volatile memory. Any error in a data block
causes all data in that block to be ignored and the data currently in the
named profile to be retained.
• Data must be input in the form of two to ten comma-separated,
NRf-format number pairs.
• The data-point number pairs in the named profile are automatically
sorted by frequency value before the profile is stored.
• When power correction is in use, the Counter uses the data points in
the currently selected profile to determine the amount of loss
correction to apply.
• If the measured frequency is between two defined data points, the
Counter uses the two defined data points to linearly interpolate the
appropriate correction value.
• If the measured frequency is above the highest frequency value in the
profile (or below the lowest frequency value), the Counter determines a
loss-correction value by using the two highest (or lowest) defined data
points to extrapolate a linear extension to the curve above the highest
data point (or below the lowest data point).
RELATED
FRONT-PANEL
KEYS
Shift + Menu (PWR CORR)