Instruction manual

Chapter 4 Command Reference
:DISPlay Subsystem
Programming Guide 4-5
4
:DISPlay Subsystem
This subsystem controls the selection and presentation of textual
information on the Counter’s display. This information includes
measurement results. :DISPlay is independent of, and does not modify,
how data is returned to the controller.
See the section titled “Programming the Counter to Display Results” on
Page 3-46 of this guide.
COMMAND :DISPlay[:WINDow]:BACKground[:STATe] . . . <Boolean>
Turns the display backlight ON or OFF.
QUERY RESPONSE Single ASCII-encoded byte, 0 or 1.
A value of 0 indicates OFF; a value of 1 indicates ON.
COMMENTS *RST: ON
Turning the display backlight OFF conserves battery power (if the
Counter has the Battery option).
If the Counter is operating from battery power (Battery option required),
the display backlight is automatically turned off after a period of
inactivity. When activity is subsequently detected (a measurable signal
is applied, a front-panel key is pressed, or a setting is changed by a
command sent over the GPIB or the RS-232 interface), the display
backlight is restored to its previous state (i.e., if the backlight was set
to OFF, it remains off; if the backlight was set to ON, it is turned on).
RELATED
FRONT-PANEL
KEYS
On/Off (Shift + Clear)