User`s guide
10
When using an E-series sensor in 438A language,
the FR (frequency) command will be accepted.
The syntax for this command is shown below.
Note: Sending this command when the active entry
channel is connected to an 8480 series sensor will
cause a syntax error.
The Agilent E-Series sensors have an upper and
lower range and are affected by the ranging com-
mands as detailed below.
• RA (auto range): The EPM series power meter
selects the most suitable range and this will
be reported in the status message range fields
for the attached sensor(s) (11 = lower range
and 12 = upper range).
• RH (range hold): The EPM series power meter will
hold on the current range irrespective of input
level and the status message range fields for the
attached sensor(s) will reflect this (01 = lower
range and 02 = upper range). The front panel
display will indicate that range hold is in effect.
• RM (manual range selection): The EPM series power
meter will select the lower range for RM1EN and
the upper range for RM2EN (RM3EN, RM4EN and
RM5EN are accepted and also select the upper
range). The status message range fields for
attached sensor(s) will reflect the range selected
(01 = lower range and 02 = upper range). The
front panel display will indicate that range hold
is in effect.
When calibrating an Agilent 8480 series sensor
using the power meter’s reference oscillator, an
error is returned if the sensor is not connected to
the power reference. This connection test does not
apply to E-series sensors.
FR
100E+03 to 999.99E+09
HZ
KZ
MZ
GZ