User`s guide

10 Specifications and Characteristics
218 Agilent N1913/1914A EPM Series Power Meters User’s Guide
Power Meter Specifications
Frequency Range
9 kHz to 110 GHz, power sensor dependant
Power Range
–70 dBm to +44 dBm (100 pW to 24 W), power sensor dependant
Power Sensors Compatibility
Agilent 8480 Series power sensors
Agilent E9300 E–Series average power sensors
Agilent E4410 E–Series average power sensors
Agilent N8480 Series power sensors
Agilent U2000 Series average USB power sensors
Single Sensor Dynamic Range
90 dB maximum (Agilent E4410 E- Series power sensors)
80 dB maximum (Agilent E9300 E- Series power sensors)
50 dB maximum (Agilent 8480 Series power sensors)
55 dBm maximum (Agilent N8480 Series power sensors)
44 dBm maximum (Agilent U2000 Series USB power sensors)
Display Units
Absolute: Watts (W) or dBm
Relative: Percent (%) or dB
Display Resolution
Selectable resolution of: 1.0, 0.1, 0.01 and 0.001 dB in logarithmic mode,
or 1, 2, 3, and 4 significant digits in linear mode
Default Resolution
0.01 dB in logarithmic mode or three digits in linear mode