User`s guide
6 Using N8480 Series Power Sensors
158 N1913/1914A EPM Series Power Meters User’s Guide
6 Connect the power sensor to the POWER REF output.
7 Press and the channel softkey to start the calibration
routine. The Calibrating pop- up is then displayed.
8 Press , . The value is displayed on the Cal Fac field.
9 Use the and keys to highlight the Cal Fac value field and press
to display the Cal Factor pop- up. Use the numeric keypad to enter
97.6 in the Cal Factor pop- up window.
10 Press
to complete the entry.
11 Connect the power sensor to the signal to be measured.
12 The corrected measurement result is displayed.
Cal
Offset
%
NOTE
When no sensor tables are selected and display mode is chosen, the
calibration factor used for the measurement is displayed in the upper window as shown in
Figure 6-7.
Single Numeric