Instruction manual

C
Copyright 2008 Agilent Technologies 
Change 6
Add TAR and change the Measurement Uncertainty value of Resistance Measurement Accuracy Test
(Grounded Device) (Page 9-20) to the following.
Resistor
Setting
Voltage
Setting
Test Limits Test Result
Measurement
Uncertainty
TAR
10
7
ohm 100 V ±0.0066 x 10
7
ohm
x 10
7
ohm ± 0.0003 × 10
7
ohm
22.6
10
11
ohm 10 V ±0.0573 x 10
11
ohm
x 10
11
ohm ± 0.0035 × 10
11
ohm
16.2