Agilent 4339B High Resistance Meter Operation Manual SERIAL NUMBERS This manual applies directly to instruments with serial number pre x JP1KD, or rmware revision 1.04. For additional important information about serial numbers, read \Serial Number" in Appendix A. Agilent Part No.
Notice The information contained in this document is subject to change without notice. This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies. Agilent Technologies Japan, Ltd.
4339B Manual Printing History March 1996 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : First Edition (part number: December 1996 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Second Edition (part number: June 1998 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Third Edition (part number: August 1998 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Fourth Edition (part numb
339B Safety Summary The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with speci c WARNINGS elsewhere in this manual may impair the protection provided by the equipment. In addition it violates safety standards of design, manufacture, and intended use of the instrument. The Agilent Technologies assumes no liability for the customer's failure to comply with these requirements.
4339B Dangerous Procedure Warnings Warnings , such as the example below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. Warning Dangerous voltages, capable of causing death, are present in this instrument. Use extreme caution when handling, testing, and adjusting this instrument.
4339B Certi cation Agilent Technologies certi es that this product met its published speci cations at the time of shipment from the factory. Agilent Technologies further certi es that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institution's calibration facility, or to the calibration facilities of other International Standards Organization members.
4339B Exclusive Remedies The remedies provided herein are buyer's sole and exclusive remedies. Agilent Technologies shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory. Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies Sales and Service O ce.
4339B Safety Symbols General de nitions of safety symbols used on equipment or in manuals are listed below. Instruction manual symbol: the product is marked with this symbol when it is necessary for the user to refer to the instruction manual. Alternating current. Direct current. On (Supply). O (Supply). This Warning sign denotes a hazard. It calls attention to a procedure, practice, condition or the like, which, if not correctly performed or adhered to, could result in injury or death to personnel.
4339B Herstellerbescheinigung GERSCHEMISSION LpA < 70 dB am Arbeitsplatz normaler Betrieb nach DIN 45635 T.
Agilent 4339B High Resistance Meter Operation Manual Manual Change Agilent Part No. N/A Oct 2012 Change 1 Add: Measurement Uncertainties (MU) on Chapter 9 Maintenance Performance Tests (Page 9-3). Measurement Uncertainties Measurement Uncertainties (MU) values shown on manual are for reference only. Always refer to test report for the latest MU values. Change 2 Add: Typical Measurement Accuracy Table remarks on Chapter 8 Supplemental Performance Characteristics.
Agilent 4339B High Resistance Meter Operation Manual Manual Change Agilent Part No. N/A Jun 2009 Change 1 Add TAR and change the Measurement Uncertainty value of Source Voltage Accuracy Test (Page 9-19) to the following. Source Voltage Setting 0V Test Limits ±0.10 V 10 V Test Result1 Measurement Uncertainty TAR V ± 0.00021V 485 ±0.12 V V ± 0.00027V 433 25 V ±0.14 V V ± 0.00054V 257 50 V ±0.18 V V ± 0.00096V 188 100 V ±0.26 V V ± 0.00185V 141 200 V ±0.42 V V ± 0.
Change 4 Add TAR and change the Measurement Uncertainty value of Current Measurement Accuracy Test (Page 9-20) to the following. Test Result1 Test Current Test Limits Measurement Time Measurement Uncertainty TAR 100 pA ±2.6 pA Long pA ± 0.27 pA 9.8 1 nA ±0.0094 nA Long nA ± 0.0007 nA 13.1 10 nA ±0.063 nA Long nA ± 0.0074 nA 8.4 100 nA ±0.53 nA Long nA ± 0.046 nA 11.4 1 A ±0.0043 A Long A ± 0.0005 μA 8.1 10 A ±0.036 A Long A ± 0.0053 μA 6.7 100 A ±0.
Change 6 Add TAR and change the Measurement Uncertainty value of Resistance Measurement Accuracy Test (Grounded Device) (Page 9-20) to the following. Resistor Setting Voltage Setting Test Limits 107 ohm 100 V 1011 ohm 10 V ±0.0066 x 107 ohm ±0.0573 x 1011 ohm C Copyright 2008 Agilent Technologies Test Result1 Measurement Uncertainty TAR x 107 ohm ± 0.0003 × 107 ohm 22.6 x 1011 ohm ± 0.0035 × 1011 ohm 16.
Agilent 4339B High Resistance Meter Operation Manual Manual Change Agilent Part No. N/A Sep 2008 Change 1 Change the Measurement Uncertainty value of Source Voltage Accuracy Test (Page 9-19) to the following. Source Voltage Setting Test Limits 0V 10 V 25 V 50 V 100 V 200 V 201 V 250 V 500 V 1000 V ±0.10 V ±0.12 V ±0.14 V ±0.18 V ±0.26 V ±0.42 V ±0.82 V ±0.90 V ±1.3 V ±2.1 V Test Result1 Measurement Uncertainty V V V V V V V V V V ± 0.000206 V ± 0.00027 V ± 0.00054V ± 0.0010 V ± 0.0019 V ± 0.
Change 4 Change the Measurement Uncertainty value of Current Measurement Accuracy Test (Page 9-20) to the following. Test Current Test Limits Measurement Time 100 pA 1 nA 10 nA 100 nA 1 A 10 A 100 A ±2.6 pA ±0.0094 nA ±0.063 nA ±0.53 nA ±0.0043 A ±0.036 A ±0.60 A Long Long Long Long Long Long Long Test Result1 pA nA nA nA A A A Measurement Uncertainty ± 0.27 pA ± 0.0007 nA ± 0.0074 nA ± 0.046 nA ± 0.0005 μA ± 0.0053 μA ± 0.
Agilent 4339B High Resistance Meter Operation Manual Manual Change Agilent Part No. N/A June 2007 Change 1 Add the following note to the “3-c. Turning ON the 4339B” (page 1-13). Note In case the setting of the ac line frequency is not match, there’s any possibility the measurement error will rise. Make sure the ac line frequency has been set up correctly. Change 2 Add the following condition to the “Measurement Accuracy” (page 8-4). 6. The ac line frequency has been set up correctly.
ᄌᦝ !"# $% &'( )*+,-.
Manual Change Agilent Part No. N/A August 2004 Printed in Japan Change Change the company name from (YOKOGAWA-) HEWLETT-PACKARD, LTD., or its abbreviation HP (YHP) to Agilent Technologies or Agilent. This document may contain references to HP (YHP) or (Yokogawa-) Hewlett-Packard. Please note that Hewlett-Packard’s former test and measurement, semiconductor products and chemical analysis businesses are now part of Agilent Technologies.
Safety Summary When you notice any of the unusual conditions listed below, immediately terminate operation and disconnect the power cable. Contact your local Agilent Technologies sales representative or authorized service company for repair of the instrument. If you continue to operate without repairing the instrument, there is a potential fire or shock hazard for the operator. n Instrument operates abnormally. n Instrument emits abnormal noise, smell, smoke or a spark-like light during the operation.
When you turn on the 4339B for the rst time, con rm the power line frequency is set correctly. 1. Press . The following menu is displayed. 2. Press until ``more'' blinks, and press . 3. Press until ``Line'' blinks, and press . A blinking item means that it is currently selected. 4. If the setting does not match the power line frequency, press between ``50 Hz'' and ``60 Hz''. Then press . to toggle the setting until ``Exit'' blinks, and press to exit this menu. 5.
4339B Contents of this Manual Chapter 1 Provides the product overview and basic measurement procedure. First time users of the 4339B should read this chapter rst. Chapter 2 Shows how to operate the 4339B from its front panel. Refer to this chapter when you wish to learn about operations using the front panel keys. Chapter 3 Describes all functions of this instrument. Refer to this chapter when you wish to learn about the functions of the front and rear panel keys and terminals.
4339B Appendix A Contains information on using the 4339Bs which were manufactured before this manual was printed. Appendix B Contains information which is required for using the handler interface. Before using the handler interface, read this appendix and set the handler interface input/output signal. Appendix C The summary of operations when the 4339B detects OVLD (Overload), Over-Current (exceeding current limit), or N.C. (No-Contact).
Contents 1. Getting Started Introduction . . . . . . . . . . . . . . . . . . . . . Overview . . . . . . . . . . . . . . . . . . . . . . Features . . . . . . . . . . . . . . . . . . . . . Accessories Available . . . . . . . . . . . . . . . . Front Panel . . . . . . . . . . . . . . . . . . . . Display . . . . . . . . . . . . . . . . . . . . . . Rear Panel . . . . . . . . . . . . . . . . . . . . Initial Inspection . . . . . . . . . . . . . . . . . . .
Turning OFF Test Voltage . . . . . . . . . . . . . . . . . . . . . . . . 2. Operating the 4339B 1-24 Introduction . . . . . . . . . . . . . . . . . . . . . . Measurement Con guration . . . . . . . . . . . . . . Selecting the Measurement Time Mode . . . . . . . . Setting the Averaging Rate . . . . . . . . . . . . . . Setting the Trigger Delay Time . . . . . . . . . . . . Setting the Parameters for Resistivity Measurement . . Entering Thickness of the DUT . . . . . . . . . . . Setting the Electrode Size .
3. Function Reference Introduction . . . . . . . . Front Panel . . . . . . . . Display . . . . . . . . . LINE Switch . . . . . . . Interlock Connector . . . UNKNOWN Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Back Space Key . . . . . . . . . . . . . . . . . . Enter Key . . . . . . . . . . . . . . . . . . . . . Minimum Key . . . . . . . . . . . . . . . . . Maximum Key . . . . . . . . . . . . . . . . Open Key . . . . . . . . . . . . . . . . . . . Calibration Key . . . . . . . . . . . . . . . . Comparator On/O Key . . . . . . . . . . . . Contact Check Key . . . . . . . . . . . . . . Display Mode Key . . . . . . . . . . . . . . . Key Lock Key . . . . . . . . . . . . . . . . . Reset Key . . . . . . . . . . . . . . .
To Set the Averaging Rate . . . . . . . . . . . . . . . . . . . . . . . . To Set Trigger Delay Time . . . . . . . . . . . . . . . . . . . . . . . . To Set the Parameters for Resistivity Measurements . . . . . . . . . . . . To Set Beeper Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . To Lock Out the Front Panel Keys . . . . . . . . . . . . . . . . . . . . To Check Contact Integrity at the Test Fixture . . . . . . . . . . . . . . . To Use the Comparator Function . . . . . . . . . . . . . . . . . .
CALCulate Subsystem . . . . . . . . . . . . . . . . . . . :CALCulate1:FORMat f REAL j SRESistivity j VRESistivity g :CALCulate1:LIMit:BEEPer:CONDition f PASS j FAIL g . . . :CALCulate1:LIMit:BEEPer[:STATe] f ON j OFF j 1 j 0 g . . . :CALCulate1:LIMit:CLEar . . . . . . . . . . . . . . . . :CALCulate1:LIMit:FAIL? . . . . . . . . . . . . . . . . . :CALCulate1:LIMit:LOWer[:DATA] . . . :CALCulate1:LIMit:LOWer:STATe f ON j OFF j 1 j 0 g . . . . :CALCulate1:LIMit:STATe f ON j OFF j 1 j 0 g . . . .
[:SENSe]:CONTact:LIMit? . . . . . . . . . . . . . . . . . . . . . . . . . [:SENSe]:CONTact:OFFSet . . . . . . . . . . . . . . . [:SENSe]:CONTact:VERify f ON j OFF j 1 j 0 g . . . . . . . . . . . . . . . [:SENSe]:CORRection:COLLect[:ACQuire] OFFset . . . . . . . . . . . . . [:SENSe]:CORRection:DATA? f OFFSet j SCAPacitance g . . . . . . . . . . [:SENSe]:CORRection[:STATe] f ON j OFF j 1 j 0 g . . . . . . . . . . . . . [:SENSe]:CURRent:APERture [MSjS] . . . . . . . . . . .
Standard Event Status Register . . . Standard Operation Status Group . . Operation Status Register . . . . . Questionable Status Register . . . . Trigger System . . . . . . . . . . . 4339B Trigger System Con guration . Idle State . . . . . . . . . . . . Initiate State . . . . . . . . . . ARM Event Detection State . . . TRIG Event Detection State . . . Sequence Operation State . . . . Data Transfer Format . . . . . . . . ASCii . . . . . . . . . . . . . . . REAL . . . . . . . . . . . . . . . Command Summary .
Display . . . . . . . . . . . . . . . . . . . Correction . . . . . . . . . . . . . . . . . . Test Sequence Program: . . . . . . . . . . . Mathematical Functions . . . . . . . . . . . Comparator . . . . . . . . . . . . . . . . . Contact Check . . . . . . . . . . . . . . . . GPIB Interface . . . . . . . . . . . . . . . Handler Interface . . . . . . . . . . . . . . Save/Recall . . . . . . . . . . . . . . . . . Continuous Memory Capability . . . . . . . . Key Lock . . . . . . . . . . . . . . . . . . General . .
Resistance Measurement Accuracy Test (Floating Device) . Resistance Measurement Accuracy Test (Grounded Device) Calculation Sheet . . . . . . . . . . . . . . . . . . . . . Source Voltage Accuracy Test . . . . . . . . . . . . . . . Ammeter Input Resistance Test . . . . . . . . . . . . . . RC Box Calibration Values . . . . . . . . . . . . . . . . Current Measurement Accuracy Test . . . . . . . . . . . Resistance Measurement Accuracy Test . . . . . . . . . . Performance Test Record . . . . . . . . . . . . . .
Figures 1-1. 1-2. 1-3. 3-1. 3-2. 3-3. 3-4. 3-5. 3-6. 3-7. 3-8. 3-9. 3-10. 4-1. 4-2. 4-3. 4-4. 4-5. 4-6. 4-7. 4-8. 5-1. 5-2. 5-3. 5-4. 5-5. 5-6. 5-7. 5-8. 5-9. 5-10. 5-11. 5-12. 6-1. 6-2. 6-3. 6-4. 6-5. 6-6. 6-7. 7-1. 7-2. 7-3. 9-1. Power Cable Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . Floating DUT Measurement Con guration . . . . . . . . . . . . . . . . . . Grounded DUT Measurement Con guration . . . . . . . . . . . . . . . . . Front Panel . . . . . . . . . . . . . . . . . . . . . . .
9-2. 9-3. 9-4. 9-5. 9-6. 9-7. 9-8. 9-9. 9-10. 9-11. 9-12. 9-13. 9-14. A-1. B-1. B-2. B-3. B-4. B-5. Ammeter O set Voltage Test Setup . . . . . . . . . . . RC Box Connection . . . . . . . . . . . . . . . . . . Current Measurement Accuracy Test Setup . . . . . . . RC Box Connection . . . . . . . . . . . . . . . . . . Resistance Measurement Accuracy Test Setup (Floating) . RC Box Connection (Floating) . . . . . . . . . . . . . . RC Box Connection for 1011 Resistor . . . . . . . . .
Tables 1-1. 2-1. 2-2. 3-1. 3-2. 3-3. 3-4. 3-5. 5-1. 5-2. 5-3. 5-4. 5-5. 5-6. 5-7. 7-1. 7-2. 7-3. 8-1. 9-1. 9-2. 9-3. 9-4. 9-5. A-1. A-2. A-3. A-4. B-1. B-2. B-3. C-1. C-2. Power Voltage Selector Setting . . . . . . . . . . . . . . Value a . . . . . . . . . . . . . . . . . . . . . . . . . Value b . . . . . . . . . . . . . . . . . . . . . . . . . Interlock Condition for 16008B and 16339A . . . . . . . . Interlock Condition for 16117B Low Noise Test Leads . . . Reset Settings . . . . . . . . . . . . . . . . .
1 Getting Started Introduction This chapter provides information to get you started using your 4339B High Resistance Meter.
Overview 4339B Overview The 4339B High Resistance Meter is designed for measuring very high resistance and related parameters of insulation materials, electronic components, and electro-mechanical components. Features Measurement Parameters: Resistance (R) Volume Resistivity (%v ) Surface Resistivity (%s ) Current (I) Test Voltage : 0.1 to 1000 Vdc R Measurement Range: 103 to 1.621016 Basic Accuracy: 0.
4339B Overview Accessories Available The following accessories are available for the 4339B: 16339A Component Test Fixture: For lead, chip, or other types of components. 16008B Resistivity Cell: For resistivity measurement of solid sheet materials. Two optional electrode sizes can be ordered in addition to the installed standard 50 mm diameter electrode.
Overview 4339B Front Panel This section gives a guided tour of the front panel. For a detailed description of each key's function, refer to Chapter 3. Each description starting with (Shift) is the secondary function of the key, which is available by pressing the blue shift key (refer to \Shift Key"). 1. Display displays measurement results, instrument states, and error messages. 2. LINE Switch turns the 4339B ON and OFF. 3.
4339B Overview 15. Trigger Mode Key selects the trigger source from Internal, Manual, or External. (Shift) Delay Key sets the trigger delay time. 16. Trigger Key triggers a measurement in the Manual trigger mode. (Shift) Sequence Abort Key cancels a measurement sequence. 17. Local Key cancels the GPIB remote state. (Shift) Address Key sets the GPIB address. 18. Recall Key recalls instrument state data from the internal memory. (Shift) Save Key stores instrument state data to the internal memory. 19.
Overview 4339B Display This section gives a guided tour of the display panel. For more detailed description of each item, refer to Chapter 3. 1. Measurement Result Display Area displays measurement result and instrument messages. 2. Measurement Setting Display Area displays instrument settings. 3. Annunciator indicates other instrument settings. Annunciator labels are as follows: a. Sequence Running indicates a measurement sequence is running. b.
4339B Overview Rear Panel This section gives a tour of the rear panel. For a more detailed description of each item, refer to Chapter 3. 1. GPIB Interface is used for controlling the 4339B from an external controller using GPIB commands. 2. Handler Interface is used to synchronize timing with an external handler. 3. Ext Trigger Terminal is used to trigger a measurement using an external signal. 4. Serial Number Plate gives the instrument's manufacturing information. 5.
Initial Inspection 4339B Initial Inspection Warning To avoid hazardous electrical shock, do not turn on the 4339B when there are signs of shipping damage to any portion of the outer enclosure (for example, covers, panel, or display) Inspect the shipping container for damage. If the shipping container or cushioning material is damaged, it should be kept until the contents of the shipment have been checked for completeness and the 4339B has been checked mechanically and electrically.
4339B Initial Inspection Instruction for Cleaning For cleaning, wipe with soft cloth that is soaked with water and wrung tightly without undue pressure.
Initial Inspection 4339B Power Cable In accordance with international safety standards, this instrument is equipped with a three-wire power cable. When connected to an appropriate ac power outlet, this cable grounds the instrument frame. The type of power cable shipped with each instrument depends on the country of destination. Refer to Figure 1-1 for the part numbers of the power cables available. Warning For protection from electrical shock, the power cable ground must not be defeated.
4339B Initial Inspection Figure 1-1.
Preparation for Use 4339B Preparation for Use Before you use the 4339B, you must set it to match the available power line voltage and frequency. Set power line voltage|refer to \Power Requirements".
4339B Preparation for Use Turning ON the 4339B 1. Connect the power cable to Power Cord Receptacle on the rear panel. 2. Push the LINE switch in. The 4339B will emit a beep and start the self test. (If any message is displayed, refer to the \Messages" in the back of this manual.) The 4339B will be ready for operation after a message similar to the one shown in the following gure is displayed. 3. When you turn on the 4339B for the rst time, con rm the power line frequency is set correctly. a. Press .
Preparation for Use Power-On Self Test 4339B When turned on, the 4339B performs self test which checks basic operation of the analog and digital circuits, display, LED, and Beeper. If an error is detected, an error message will be displayed on the LCD. For more information, refer to \Con guration Key " in Chapter 3. Warning Before turning the 4339B OFF, the High Voltage indicator must be OFF. When the High Voltage indicator is ON, wait until the High Voltage indicator is OFF.
4339B Using Front Panel Keys Using Front Panel Keys The 4339B has three types of keys as follows: Direct Execution Type Keys Toggle Type Keys Selection Type Keys Value Setup Type Keys Note If you want to exit an operation and go back to the measurement mode, press several times until you are back to the measurement mode. Direct Execution Type Keys Pressing a direct execution key performs the pressed key's function immediately.
Using Front Panel Keys 4339B Value Setup Type Keys Pressing a value setup type key will display the parameter entry display and prompt you to enter a value. For example, pressing displays, The displayed value can be entered/changed by using one of the following: Numeric Keys Maximum and Minimum Keys Down and Up Arrow Keys Back Space Key Value Setup Using Numeric Keys Some commands use numeric parameters. Fore example, to set the test voltage value to 100 V, 1. Press . 2. Press .
4339B Value Setup Using Maximum and Minimum Keys Using Front Panel Keys These keys enter the maximum and minimum numeric value in place of having to use the numeric keys. For example, 1. Press . The maximum value of source voltage, 1000 V is displayed. 2. Press . The minimum value of source voltage, 0.0 V is displayed. 3. Press . Value Setup Using Down and Up Arrow Keys Increases or decreases the numeric entry. For example, 1. Press to select current limit menu. 2.
Basic Operation 4339B Basic Operation This section provides the basic measurement procedure for the 4339B. Follow the instructions and become familiar with the 4339B's operation.
4339B Connecting Test Fixture Connecting Test Fixture The 4339B has the capability to measure either oating or grounded DUTs. The measurement con guration is di erent for oating and grounded DUT measurements, and a Shunt connector is used to change the measurement con guration. Refer to the Operation and Service Manual of the test xture for more information on how to connect it. Warning Do NOT touch the UNKNOWN terminals or the electrodes of the accessory.
Performing Calibration 4339B Resetting 4339B Resetting the 4339B changes its settings to their default state. 1. Press to select the reset menu. 2. Press to select Yes (Yes is blinking), and press . For more information about the default state of the 4339B, refer to \Reset Key Chapter 3. " in Performing Calibration Calibration cancels measurement errors due to environmental temperature change. 1. Press . The 4339B displays the calibration menu. 2. Press or until ExecCal blinks, and press .
4339B Performing OPEN Correction Setting Test Voltage The 4339B has an internal DC voltage source. To set voltage level of this source: 1. Press to display the voltage setup menu. 2. Enter the value using the numeric keys, and press .
Performing OPEN Correction 4339B Performing OPEN Correction The OPEN correction function cancels measurement errors caused by residuals and stray capacitance of the test xture, and stores data for the contact check function. 1. Separate each electrode of the test xture, con rming that nothing is connected to the electrodes. For details about xture operation, refer to each xture's Operation and Service Manual. 2. Press .
4339B Selecting Measurement Range Selecting Measurement Parameter The 4339B has four measurement parameters. DC resistance: R DC current: I Surface Resistivity: Rs Volume Resistivity: Rv Select the measurement parameter as follows: 1. Press . The following menu appears. 2. Press or parameter. Note until your desired parameter blinks.
Turning OFF Test Voltage 4339B 2. Press or until the desired range is displayed. Or, input the current value to be measured, and let the 4339B select the optimum measurement range setting. . 3. press Applying Test Voltage Press . The source voltage speci ed using the key is applied immediately, and the V Output indicator turns ON. The displayed value is the valid measurement result when the applied source voltage is applied. Warning Pressing may output dangerous voltage level up to 1000 Vdc.
2 Operating the 4339B Introduction This chapter provides step-by-step instructions for using the 4339B High Resistance Meter. It includes the following sections: Measurement Con guration Making a Measurement Testing the 4339B If You Have a Problem Refer to Chapter 3 for the description of each front panel key function.
Measurement Con guration Measurement Con guration 4339B This section discusses the 4339B's general con guration topics that apply to all measurement functions. Selecting the Measurement Time Mode The current measurement time mode setting (Short, Medium or Long) is indicated by the annunciator( 9 ). To select the measurement time mode: Press until the measurement time mode is set to the desired mode. Setting the Averaging Rate To set the averaging time: 1. Press . 2.
4339B Measurement Con guration Setting the Parameters for Resistivity Measurement To set the parameters to calculate the resistivities (DUT thickness, Electrode Size) or to show the current settings: 1. Press . The resistivity parameter setup main menu is displayed. Entering Thickness of the DUT To enter the thickness of the DUT or to show the current setting: 2. Select Thickness using or and press . 3. Enter the desired value, or leave the current setting. Press 4.
Measurement Con guration 4339B When You Use Your Own Resistivity Cell: a. Select User using or and press . b. To set the e ective area value, select Area using or Enter the value using the numeric keys, and press Enter the value using the numeric keys, and press or 6. Select Exit using menu. 2-4 Operating the 4339B or or and press and press or and press to enter the value and to exit. and press Enter the value using the numeric keys, and press e. To exit, select Exit using .
4339B Making a Measurement Making a Measurement Triggering a Measurement The 4339B has four trigger source modes: Internal, Manual, External, or Bus. The Trigger annunciator( 9 ) shows which trigger source is selected. Note When the bus trigger mode is selected, none of the Trigger annunciators( 9 ) are ON. The bus trigger mode can be set by GPIB commands only. To Trigger Internally 1. Press until the Int trigger annunciator( 9 ) is ON. To Trigger Manually 1.
Making a Measurement 3. Press . 4. Enter the upper limit value using the numeric keys, and press Note 4339B . When you want to set the limits only for the lower (or upper) limit, set the upper (lower) limit value to the maximum (minimum) value by using ( ) in the limit value entry menu. To start sorting: Press . To display the sorting results: Press . The following menu appears. Select Comprtr using or and press .
4339B Making a Measurement Displaying Deviation Data The 1 before the measurement parameter on the LCD display indicates that the displayed value for the parameter is a deviation value. Setting the Reference Value A reference value is needed for a deviation measurement. To set the reference value or to show the current setting: 1. Press . The following menu appears. 2. Select 1RefEnt using or and press .
Making a Measurement 4339B Selecting the Deviation Display Mode To select the deviation display mode: 3. Select ModeSet using or and press . The following menu appears. The 1ABS mode displays the di erence between the measured value and the reference value. To select the 1ABS mode: Select 1ABS using or and press . The 1% mode displays the the di erence between the measured value and the reference value as a percentage of the reference value. To select the 1% mode: Select 1% using or and press .
4339B Making a Measurement Using the Measurement Sequence Function The measurement sequence function allows you to control a series of measurements in a sequence. Refer to \Measurement Sequence Mode Key " in Chapter 3 for the details of the measurement sequence function. Selecting the Measurement Sequence Mode The 4339B has two sequence modes: the (Single) mode and the (Continuous) mode. The current sequence mode setting can be displayed in the Measurement Settings area of the LCD display.
Making a Measurement 4339B To Set Interval Time and Number of Repetitions To set the measurement interval time and the number of measurement points (equivalent to Memory Size) used in the continuous mode: 1. Select Intvl using or and press . 2. Enter the interval time using the numeric keys, and press Then the Memory Size entry menu appears. 3. Enter the number of measurement points, and press . .
4339B Making a Measurement 5. Select Up (Count up mode) or Down (Count down mode) using , and press . 6. If you want make the 4339B to emit a beep when the measurement sequence is furnished, select Beep using or and press . Then select On using or and press . Starting Measurement Sequence Press ON. Warning . The measurement sequence is started and the Seq Running annunciator( 9 ) turns Pressing may cause the 4339B to output dangerous voltage levels up to 1000 Vdc.
Making a Measurement 4339B Setting Contact Check The Cont Chk annunciator( 9 ) indicates the current contact check status. Note Before performing the contact check, perform an OPEN correction to measure the reference values for the contact check. Refer to \ Correction" in Chapter 1. Performing OPEN To enable or disable the contact check function: 1. Press . 2. Select ON/OFF using or 3. Select On or Off using 4. Select Exit using and press or or , and press , and press to select.
4339B Making a Measurement Contact check compares a measured capacitance data of dut with a limit data. Interpretation of the limit data and the measured capacitance data of the dut are shown below. Reading the Limit Data of Contact Check Function 1. Press . 2. Select more using or 3. Select LimitC using , then press or . , then press . 4. The limit value is displayed. The limit data (Limit) is calculated by the following equation.
Making a Measurement 3. Select MeasC using 4339B or , then press . 4. The capacitance value at measured DUT is displayed. The condition of N.C.(No Connection) is shown below. MeasC < Limit Where MeasC Limit Data displayed in MeasC Data displayed in Limit Setting Current Limit To set the current limit: 1. Press . 2. Enter the value using the numeric keys, and press . Setting Current Monitor To turn ON the current monitor, press Select On using or and press . The following menu is dislayed. .
4339B 2. Select Beep using Making a Measurement or and press 3. Select the desired Beep mode using 4. To exit, select Exit using or . or , and press and press . .
Making a Measurement 4339B Setting Display Mode and Display Format If you press , the display mode selection menu appears. Select the desired mode using or and press (Data: displays the measurement data, Comprtr: displays the comparator results, Off: turns the display OFF). If you select Formt, the measurement data display format menu appears. If you select Digit, the display digits selection menu appears. Select the desired display digits using or , and press .
4339B Making a Measurement Changing Measurement Settings Display Mode Each time is pressed, the current settings of the 4339B are displayed one after another on the right side of the LCD display. The information displayed and the order of display is as follows: 1. Test voltage setting Vout and Current limit Clmt 2. Averaging rate AVG and Trigger delay time DLY 3. Sequence mode SEQ and Charge time CHRG 4. Sequence interval time INTV and Memory size MEM 5. Comparator limits Hi, Lo 6.
Making a Measurement 4339B Saving and Recalling Instrument Settings The 4339B can save and recall the instrument's settings into non-volatile memory (EEPROM). To save the current settings: 1. Press . 2. Enter the register number (0 to 9) into which you want to save the settings. 3. Press Note to save. Record the register number you used for future reference. To recall a setting. 1. Press . 2. Enter the register number (0 to 9) from which you want to recall the settings. 3. Press to recall.
4339B Making a Measurement Setting the GPIB Address 1. Press . 2. Enter the desired value, then press to set the value and to exit. When you just want to see the current setting, press . Printing Measurement Data The 4339B can print measurement data to an GPIB compatible printer without using an external controller. To do so, you must: 1. Set the printer to listen-always mode. (Consult the printer manual for instructions.) 2. Connect the printer to the 4339B's GPIB port. 3. Turn the printer ON. 4.
Testing the 4339B 4339B Testing the 4339B Performing a Self-Test The 4339B has a self-test function to check its basic performance. 1. Press . 2. Select more using or and press 3. Select Test using or and press is displayed, refer to \Error Messages." 4. Select Exit and press to exit. 2-20 Operating the 4339B . to execute the self test.
4339B Testing the 4339B Testing the Front Panel Key's Functionality The 4339B has a service function to test the functionality of the front panel keys which this section describes. (There are another service functions which test the handler interface | refer to Chapter 9, and the ROM, RAM , and EEPROM | for use by service personnel only.) 1. Press . 2. Select more using or and press . 3. Select SVC using or and press . 4. Select KEY using or and press . 5.
If You Have a Problem 4339B If You Have a Problem If the Display is Blank and the 4339B Appears Dead If the display is blank, and even the annunciators are not ON: Check the fuse. If an Error Message is Displayed Refer to \Messages." If the 4339B does not Accept Any Key Input Check whether the Rmt annunciator( 9 ) is ON. Check whether the external controller is disabling all the front-panel controls using the LOCAL LOCKOUT command. If so, send the LOCAL command from the external controller. Press .
3 Function Reference Introduction This chapter provides information on all the 4339B's functions.
Front Panel 4339B Front Panel Note Figure 3-1. Front Panel In this manual, the blue shift key is expressed as the key is not labeled with the word \blue". , even though the top of Display The display serves two functions | character display and annunciator display. The character display shows the measurement result, instrument setting information, and instrument messages. The 4339B has four measurement display modes. For details about the display modes, refer to \Display Mode Key ".
4339B Front Panel LINE Switch The LINE Switch turns the 4339B ON or OFF. In the 1 (ON) position power is applied and all operating voltages are applied to the instrument. In the 0 (OFF) position no power is applied and no operating voltages are applied to the instrument. The 4339B's settings are held in backup memory for about 72 hours after power is turned OFF. Refer to \Reset Key " for the backed up settings. Note V Output ON /OFF state is not saved, and the 4339B sets the V Output to OFF at power-on.
Front Panel 4339B UNKNOWN Terminals These are terminals used to connect the test xture or test leads for measuring a DUT. Input Is connected to the DUT. The connector type is triaxial, which is con gured as a center conductor, an inner shield, and an outer shield. The outer shield is connected to ground, and shields the test signal from noise. The inner shied, to which nothing is connected, guards the very low current signal, and cuts the leakage current of the contact check signal.
4339B Front Panel V Output Indicator The V output indicator indicates that the test voltage is applied to the DUT. Source Voltage key The source voltage key sets the test voltage value from 0 to 1000 V. Resolution of the test voltage is as follows: Voltage Range 0 V Test Voltage 200 V 200 V < Test Voltage 1000 V Resolution 0.1 V 1V Current Limit Key The current limit key sets the limitation value of current owing through the DUT.
Front Panel 4339B Measurement Sequence Mode Key The Measurement Sequence Mode key toggles the measurement sequence mode ON and OFF. This feature provides an automatic measurement process. The 4339B has two sequence modes: the single mode and the continuous mode. Single Mode Figure 3-2. Sequence Mode Figure 3-2(a) shows a measurement sequence in the Single mode. The single mode is appropriate for measuring a DUT when the operator de nes the charge time (for example, 1 minute).
4339B Front Panel 1. At the Ready state, pressing starts a measurement sequence, and the test voltage is turned ON. 2. The 4339B waits for the de ned Charge Time while the applied voltage is charging the DUT. 3. After that, the 4339B repeats measuring the DUT as many times as is de ned by the Memory Size with a de ned Interval Time between measurements.
Front Panel Time Display Function 4339B The time display function displays the elapsed time during the measurement sequence is running. The following menu appears when TimeDisp is selected: Mode | sets the time display mode. Sec Displays the elapsed time in seconds. % Displays the number of measurement completed points (in Count up mode) or remaining measurement points (in Count down mode) as a percentage of the total number of measurement points. O Turns OFF the time display function.
4339B Front Panel Parameter Description D1 Main electrode diameter D2 Guard ring diameter B E ective area coe cient Available Value 26 mm1 50 mm1 76 mm1 38 mm1 70 mm1 88 mm1 0 B 1 (0.0001 resolution) 1 These parameter's values are xed and cannot be changed.
Front Panel 4339B Show Setting Key The Show Setting Key changes the mode of the Measurement Settings display on the right side of the LCD display. Each time is pressed, the display mode is changed as follows: 1. 2. 3. 4. 5. 6.
4339B Front Panel Trigger Key The trigger key triggers a measurement when the 4339B is in the Manual trigger mode. Refer to \Trigger Mode Key " for more information. Warning When the measurement sequence mode is ON, pressing may cause the 4339B to output dangerous voltage levels up to 1000 Vdc. Do NOT touch the UNKNOWN terminals or the electrodes of the accessory when the V Output indicator is ON. Sequence Abort Key The Sequence Abort key aborts the running measurement sequence program.
Front Panel 4339B Trigger Mode Key The Trigger Mode key selects the trigger source. The available trigger sources are as follows: Int (Internal) Trigger Mode. The 4339B is triggered automatically and continuously. (Default) Man (Manual) The 4339B is triggered when is pressed. Ext Bus (External) The 4339B is triggered by a pulse input through the External Trigger terminal or the handler interface. Refer to \External Trigger". (Available only in GPIB remote mode.
4339B Front Panel Left/Down and Right/Up Arrow Keys These keys serve the following functions: When you enter a value for the value setup keys, these keys increase or decrease the setting value. When you select an item for the selection keys, these keys are user to activate the item (the selected item blinks). 0, ... , 9, .(point), 0(minus) Keys ... These keys are used to enter numeric data into the 4339B. Pressing data input.
Front Panel 4339B Open Key The Open key is execute the OPEN correction measurement to obtain the OPEN correction data. The 4339B uses this data to cancel the leakage of the test xture and to obtain the limit value for the contact check function. The correction data is stored in non-volatile memory (EEPROM). See \ If Performing OPEN Correction" in Chapter 1 for more information. is pressed, the following OPEN correction menu is displayed. OpenMeas MeasVal Exit Peroforms the OPEN correction.
4339B Front Panel Contact Check Key The Contact Check key toggles the contact check function ON and OFF. This function monitors whether the DUT is properly connected to the test xture or test leads. When the contact check fails, N.C. (No-Contact) will be displayed on the LCD display. The measurement status of the GPIB output data is set to 2 (No-Contact). Refer to \Address Key " for the GPIB output data. The /NO CONTACT pin of the handler interface is asserted.
Front Panel 4339B Display Mode Key The Display Mode key selects the display mode from the following choices: Data Displays the measurement result. (default) When the contact check function is ON and the contact check has failed, N.C. (no-contact) will be displayed instead of the measurement data. Comprtr (When comparator function is ON,) Displays the result of the comparison as HIGH (greater than upper limit), IN (Passed), and LOW (less than lower limit), or N.C. (contact check failed).
4339B Front Panel Reset Key The Reset key resets all instrument settings and correction data to the default values. The 4339B can also be reset by sending GPIB commands :SYST:PRES and *RST, with the same results, except in a couple of cases. The following table lists the di erence among the results of pressing the Reset key and sending the :SYST:PRES and *RST command, and also lists where the settings are stored in. Data stored in back-up memory is held for about 72 hours after powering OFF.
Front Panel 4339B Table 3-3. Reset Settings (continued) Item Reset key ON Beep ON/OFF state Beep mode FAIL mode ASCII Data transfer format Power line frequency No e ect No e ect GPIB Address Key lock N/A :INT:CONT ON SYST:PRES1 *RST1 No e ect OFF OFF Stored in EEPROM EEPROM EEPROM EEPROM EEPROM None None Con guration Key The Con guration key allows you to set the o set-error canceling, the beep mode, and the power line frequency, and to run the self-test.
4339B Front Panel Beep Comparison results output to the beeper are classi ed into following modes: OFF Does not emit a beep (no output to beeper). PASS Emits a beep when the comparison result is IN. FAIL Emits a beep when the comparison result is HIGH, LOW or N.C.. (default) Note In the OFF mode, the 4339B does not emit a beep when a system error or operation error occurs. more Displays other items of this menu. Line | Available power LINE frequency selections are 50 Hz and 60 Hz.
Rear Panel 4339B Rear Panel Figure 3-4. Rear Panel External Trigger The Ext (External) Trigger terminal is used to trigger the 4339B by inputting a positive-going TTL pulse, when the 4339B is set to external trigger mode. Figure 3-5 shows the speci cations required for the TTL pulse. Figure 3-5.
4339B Rear Panel LINE Fuse Holder The 4339B's line fuse is selected depending on the LINE Voltage selection. Refer to Table 1-1. LINE Voltage Selector The Line Voltage Selector is used to match the 4339B to the power line voltage being used. Refer to Table 1-1. Power Cord Receptacle The Power Cord Receptacle is used to plug in the power cord. Power Cord The 4339B is equipped with a three-conductor power cord that, when plugged into the appropriate AC power receptacle, grounds the instrument.
Rear Panel 4339B Handler Interface Handler interface is used to synchronize timing with an external handler. Before using the handler interface, you must connect pull-up resisters to enable the output signals and set the dip switch to select the voltage level to match the input signals. Refer to Appendix B for these procedures.
4339B Rear Panel Table 3-4. Contact Assignment for Comparator Function Description Signal Name1 Pin No. 1 2 EXT DCV1 EXT DCV1 External DC Voltage 1: DC voltage supply pins for DC isolated open collector outputs (/HI, /IN/, /LO/, /NO CONTACT). Maximum voltage is +24 V, minimum +5 17 /HV OFF Voltage Source OFF: When this line is asserted, output of the voltage source is disabled. 18 /KEY LOCK Key Lock: When this line is asserted, all of the 4339B's front panel key functions are disabled.
Rear Panel 4339B Figure 3-7. Timing Diagram Note This timing diagram is also applied when the contact check fails, because the measurement is performed and the measurement result is output, even if the contact check fails. Note All the output signals are available only when the comparator function is ON.
4339B Rear Panel GPIB Interface The GPIB Interface is used for remote control of the 4339B using the General Purpose Interface Bus (gpib). GPIB is a standard for interfacing instruments to computers, and supports for IEEE 488.1 , IEC-625 , IEEE 488.2 , and JIS-C1901 . GPIB allows instruments to be controlled by an external computer which sends commands or instructions to and receives data from the instrument.
Theory of Operation 4339B Theory of Operation This section provides the 4339B's theory of operation. In \Overall Measurement Theory" and \Overall Block Diagram", we will discuss the measurement theory with the ungrounded DUT measurement con guration. These descriptions can also be applied to the measurement of a grounded DUT, considering the test signal ow shown in Figure 3-10 (b). Overall Measurement Theory The 4339B measures the Device Under Test (DUT) resistance in the following manner. Figure 3-8.
4339B Theory of Operation Overall Block Diagram Figure 3-9 shows the overall block diagram of the 4339B. Figure 3-9. 4339B Overall Block Diagram The voltage source applies a dc voltage which is set by the keyboard or by an GPIB command. The I-V converter, which is directly connected to the input, converts the current into a voltage ramp, of which the slope is proportional to the input current.
Theory of Operation 4339B Grounded and Ungrounded DUT Measurement Con guration The 4339B has the capability to measure both grounded and ungrounded DUTs. Figure 3-10. Ungrounded and Grounded DUT Measurement In Figure 3-10 (a), the Ground and the Guard terminal are connected by a shut connector. The Input terminal is connected to one side of the DUT, while the other side of the DUT is connected to the Output terminal.
4 Remote Operation Introduction This chapter provides step-by-step instructions for controlling the 4339B using GPIB remote mode. The examples in this manual use the HP 9000 series 200 or 300 BASIC language. This chapter covers the following: Getting started Setting up the 4339B Triggering a measurement Retrieving measurement data Other features If you have a problem Refer to Chapter 5 for the description of each GPIB command. Note In this chapter, all commands are spelled in abbreviated (short) form.
Getting Started 4339B Getting Started This section will teach you the basics of operating the 4339B in GPIB remote mode (from now on referred to as remote). This includes reading the GPIB address, sending commands to the 4339B, and retrieving data from the 4339B. Input/Output Statements The statements used to operate the 4339B in remote depend on the computer and the programming language being used. In particular, you need to know the statements the language uses to input and output information.
4339B Getting Started Query Commands There are several commands in the alphabetic command directory that end with a question mark. These commands are called query commands because each returns a response to a particular question. In addition to the queries described above, you can create others by appending a question mark to most commands. Getting Data from the 4339B The 4339B can output readings and responses to query commands.
Getting Started 4339B To Control the 4339B from an External Computer Most measurements can be modeled by the following simple four step sequence: 1. Set up the instrument. Typically, you begin the setup step by sending the *RST command to set the instrument to its default settings. Next, if you need values di erent from the default settings, change the settings one by one as required. 2. Trigger the measurement.
4339B Getting Started To Set Up the 4339B To Reset the 4339B The following commands reset the 4339B: *RST :SYST:PRES Note The *RST command also initiates the trigger system also and places trigger sequence in the IDLE state. Refer to \Trigger System" in Chapter 5 for more information.
Getting Started 4339B To Apply the Test Voltage The following commands applies the test voltage: :OUTP For example, OUTPUT 717;":OUTP ON" : OUTPUT 717;":OUTP OFF" Warning Sending :OUTP command may cause the 4339B to output a dangerous voltage levels up to 1000 Vdc. Do NOT touch the UNKNOWN terminals or the electrodes of the accessory, when the V Output indicator is ON. To Set the Current Limit The following command sets the current limit value: :SOUR:CURR:LIM For example, to set 0.
4339B Getting Started To Perform OPEN Correction The following command performs an OPEN correction: :SENS:CORR:COLL OFFS For example, : DISP "Connect the test fixture without a DUT, then press `Continue'." PAUSE OUTPUT 717;":SENS:FUNC 'CURR'" OUTPUT 717;":SENS:CURR:APER 0.39" OUTPUT 717;":SOUR:VOLT 10" OUTPUT 717;":OUTP ON" REPEAT OUTPUT 717;":FETC?" ENTER 717;S,D UNTIL ABS(D)<5.
Getting Started To Set the Parameters for Resistivity Measurements 4339B The following commands set the parameters for resistivity measurements: :CALC:RES:STH :CALC:RES:EPER :CALC:RES:GLEN :CALC:RES:EAR For example, to measure the volume resistivity, when the thickness of DUT is 2 mm, and when you use the 50 mm electrode and 70 mm guard ring of the 16008B Resistivity Cell, that is, Thickness is 0.002 m E ective Area is 0.0019635 (= 2 (0.05/2)2 ) m2 E ective Perimeter is 0.1885 (= 2 (0.05+0.
4339B Getting Started To Use the Comparator Function The following commands control the comparator function: :CALC:LIM:CLE :CALC:LIM:FAIL? :CALC:LIM:LOW :CALC:LIM:LOW:STAT :CALC:LIM:UPP :CALC:LIM:UPP:STAT :CALC:LIM:STAT For example, to set the limit values and to enable the comparator function, : OUTPUT 717;":CALC:LIM:LOW 1E10" OUTPUT 717;":CALC:LIM:UPP 1E11" OUTPUT 717;":CALC:LIM:STAT ON" OUTPUT 717;":FETC?" ENTER 717;S,D,C : To Display a Deviation Measurement The following commands display deviation r
Getting Started For example, to wait until the OPEN correction is completed, OUTPUT 717;":SENS:CORR:COLL STAN1" Perform OPEN correction OUTPUT 717;"*OPC?" Wait for OPEN correction completed ENTER 717;A To Get the Current Instrument Settings The following command gets the current instrument settings: *LRN? For example, DIM A$[1000] : OUTPUT 717;"*LRN?" ENTER 717;A$ PRINT A$ To Save and Recall Instrument Settings The following commands save and recall instrument settings: *SAV *RCL To save the instrument s
4339B To Trigger a Measurement To Trigger a Measurement The following commands are used to trigger measurements from an external controller and to handle the 4339B's trigger system. Refer to \Trigger System" in Chapter 5 for the information about the 4339B's trigger system. :TRIG :TRIG:SOUR :INIT :INIT:CONT :ABOR *TRG :FETC? Group Execution Trigger (GET) Follow the procedure below to perform successive measurements automatically (initial setup). 1.
Waiting For Completion Of Measurement (detecting completion of measurement) 4339B Waiting For Completion Of Measurement (detecting completion of measurement) You can detect the status of the 4339B by using the status register, as described in this section. For information on the entire status report system (for example, information on each bit of the status register), refer to Chapter 5, \GPIB Reference." The measurement state is indicated by the operation status register (refer to Table 5-5).
4339B Waiting For Completion Of Measurement (detecting completion of measurement) Sample program Figure 4-3 shows a sample program to detect the completion of measurement using an SRQ. This program stops the trigger system, sets up SRQ, and then starts up the trigger system once. When an SRQ of the completion of the measurement occurs, it displays a \Measurement Complete" message and nishes. The program is detailed below. Line 20 Sets the GPIB address.
Reading Out Measured Result 4339B Reading Out Measured Result This section describes how to read out the measured result. You can read out the measured result in two ways: reading out data for each measurement or reading out data of several measurements in batch. The three commands shown in the table below can be used to read out the measured data for each measurement.
4339B Reading Out Measured Result Reading out measured result using *TRG command This command actually performs two tasks: it triggers the instrument and returns the results. It is useful, for example, when you want to retrieve measurement results immediately after triggering the instrument from an external controller. The readout procedure using the *TRG command is described below. 1. Use the :TRIG:SOUR command to set up the trigger mode to the GPIB trigger (Bus). 2. Execute the *TRG command. 3.
Reading Out Measured Result 4339B Figure 4-4.
4339B Reading Out Measured Result The program of Figure 4-5 is detailed below. Lines 50 to 60 Sets the GPIB address. Line 70 Sets the data transfer format to the binary format. Lines 90 to 100 Turns ON the trigger system continuous startup and sets the trigger mode to the GPIB trigger. Lines 120 to 130 Reads out the ON/OFF state of the comparator function and assigns it to the Comp ag variable. Lines 170 to 210 After the trigger system state transitions to the trigger wait state, triggers the instrument.
Reading Out Measured Result 4339B Figure 4-5.
4339B Reading Out Measured Result Reading out measured result using :FETC? command You can use this readout method when you want to trigger the instrument from any source other than an external controller or when you want to perform a process that is between triggering the instrument and reading out the measured result. The readout procedure using the :FETC? command is described below. 1. Set up the trigger mode as necessary. 2. Trigger the instrument by using the method for the trigger mode.
Reading Out Measured Result 4339B Figure 4-6.
4339B Reading Out Measured Result The program of Figure 4-7 is detailed below. Lines 50 to 60 Sets the GPIB address. Line 70 Sets the data transfer format to the binary format. Lines 90 to 100 Turns ON the trigger system continuous startup and sets the trigger mode to the external trigger. Lines 120 to 130 Reads out the ON/OFF state of the comparator function and assigns it to the Comp ag variable.
Reading Out Measured Result 4339B Figure 4-7.
4339B Reading Out Measured Result Data Retrieval After triggering, the output format is, ,, Where, Measurement status 0 : Normal 1 : Overload 2 : No-Contact 4 : Over current (exceeding current limit) Measurement data Comparison result (no output when the comparator function is OFF) 1 : In 2 : High 4 : Low 8 : No-Contact Remote Operation 4-23
To Retrieve Data E ciently 4339B To Retrieve Data E ciently The basic procedure to retrieve measurement data is described in \To Trigger a Measurement". This section describes how to retrieve the measurement data e ciently. To Use Data Bu er You can use the data bu er to temporarily store the results of several measurements and then later read out these results in batch.
4339B To Retrieve Data E ciently To Perform a Measurement Sequence The following commands perform the sequence measurement: :ARM:SOUR :ARM:DEL :TRIG:SOUR :TRIG:TIM :TRIG:COUN Warning When the 4339B is in the sequence measurement mode, triggering a measurement may output a dangerous voltage levels up to 1000 Vdc. Do not touch the UNKNOWN terminals or the electrodes of the accessory, when the V output indicator is ON.
Other Features 4339B Other Features To Test the 4339B The following command runs the 4339B's internal self test: *TST? For example, : OUTPUT 717;"*TST?" ENTER 717;A IF A<>0 THEN PRINT "Self Test: Error" IF BIT(A,0) THEN PRINT "RAM Test Error" IF BIT(A,1) THEN PRINT "EPROM Test Error" IF BIT(A,2) THEN PRINT "Calibration Data Test Error" IF BIT(A,3) THEN PRINT "User's Data Test Error" IF BIT(A,4) THEN PRINT "A/D Converter Test Error" IF BIT(A,5) THEN PRINT "Backup Memory Test Error" : To Read the Error Que
4339B Other Features For example, to detect the measurement completion: OUTPUT 717;":SYST:OPER:ENAB 16" Enable Measurement bit of Operation Status Register OUTPUT 717;"*SRE 128" Enable Operation Status Register Summary bit REPEAT Wait until the Operation Status Register Summary A=SPOLL(717) bit is set UNTIL BIT(A,7) For example, to generate an interrupt when an error occurs in the 4339B: : DIM Err$[50] ! : OUTPUT 717;"*CLS" Clears status byte register OUTPUT 717;"*ESE 48" Sets Command Error Bit and Execut
Sample Program 4339B Sample Program The following sample program measures resistivity of insulation lm using the 16008B Resistivity Cell with 50 mm main electrode. Warning Do NOT touch the UNKNOWN terminals or the electrodes of the accessory, when the High Voltage indicator is ON, the 4339B outputs dangerous voltage levels up to 1000 Vdc. Before handling the 4339B or the accessory, con rm that the High Voltage indicator is OFF.
4339B 400 410 420 430 440 450 460 470 480 490 500 510 520 530 540 550 560 570 580 590 600 Sample Program OUTPUT @Hp4339;":CALC:RES:EAR 0.0019635" OUTPUT @Hp4339;":CALC:RES:EPER 0.18850" OUTPUT @Hp4339;":CALC:RES:GLEN 0.01" ! OUTPUT @Hp4339;":SENS:FUNC 'RES'" DISP "Turn the Volume/Surface selector to `Volume'." PAUSE DISP OUTPUT @Hp4339;":CALC:FORM VRES" OUTPUT @Hp4339;"*TRG" ENTER @Hp4339;S,D PRINT "Volume Resistivity:",D;"[OHMcm]","Status:";S ! DISP "Turn the Volume/Surface selector to `Surface'.
If You Have a Problem If You Have a Problem 4339B Check all GPIB addresses and connections; most GPIB problems are caused by an incorrect address or bad or loose GPIB cables.
5 GPIB Reference Introduction This chapter provides the information needed to control the 4339B through the GPIB interface in the GPIB Remote mode. GPIB Command Reference Status Reporting Trigger System Data Transfer Format GPIB Commands The 4339B's GPIB commands are compatible with the Standard Commands for Programmable Instruments (ACPI ). SCPI is the instrument command language for controlling instruments that go beyond IEEE 488.
GPIB Commands 4339B Colon (:) When a colon is placed between two command mnemonics, the colon moves the current path down one level on the command tree. When the colon is the rst character of a command, it speci es the following command mnemonic is a root-level command. Semicolon (;) A semicolon separates two commands in the same message without changing the current path. Figure 5-1 shows examples of how to use the colon and semicolon to navigate e ciently through the command tree.
4339B GPIB Commands Case Letter cases (upper and lower) are ignored. Program Message Terminator Program message must be ended with one of the three program message terminators, , <^END>, or <^END>. <^END> means that End Of Identify (EOI) is asserted on the GPIB interface at the same time the preceding data byte is sent. For example, the HP BASIC OUTPUT statement is automatically sent after the last data byte.
GPIB Commands 4339B The subsystem commands use extended numeric parameters. Extended numeric parameters accept all numeric parameter values and other special values, for instance, MAXimum, MINimum or UP, DOWN. The special values available are described in the command's reference description. Query response of is always a numeric value in (integer) or ( oating point) format.
4339B GPIB Commands Query and Response Message Syntax All subsystem commands can be queried except for the commands described as \no query" in the command reference. To send a query message, add a question mark, ? after the last command mnemonic. :AVER:COUN? A response message may contain both commas and semicolons as separators. When a single query command returns multiple values, a comma is used to separate each data item.
Command Reference 4339B Command Reference Notations In this chapter, each subsystem is documented using the tabular format instead of the command tree as follows. The following conventions and de nitions are used to describe GPIB operation. COMMAND FORMat [:DATA] COMMAND INITiate :CONTinuous [:IMMediate] PARAMETER fASCiijREAL[,64]g PARAMETER Commands closest to the root-level are at the top of the table. Lower nodes in the hierarchy are indented one position to the right.
4339B ABORt Command ABORt Command :ABORt Resets the trigger system and places all trigger sequences in Idle state. Any actions related to the trigger system that are in progress, such as acquiring a measurement, are aborted immediately. The execution of an :ABORt command will set any pending operation ag to FALSE, for example ags that were set by the initiation of the trigger system. Refer to \Trigger System". Unlike *RST, :ABORt does not alter the settings programmed by other commands.
4339B ARM Subsystem The 4339B has the measurement sequence feature which provides a measurement sequence (charging DUT, and then measuring it several time with interval) by a single operation | triggering a measurement. The ARM subsystem controls the sequence measurement feature, working with the TRIGger subsystem. The purpose of the ARM subsystem is to qualify an ARMing sequence, before enabling the TRIGger system, to charge the DUT before measuring it.
4339B ARM Subsystem :ARM[:SEQuence1][:LAYer]:DELay [MSjS] Sets the DUT charging time before a DUT is measured. The charge time de nes the waiting time when the 4339B triggered a measurement to when the measurement will actually start. The test voltage is applied during this time.
4339B CALCulate Subsystem The CALCulate subsystem controls measurement data processing listed below. 1. To select measurement parameter (CALCulate1:FORMat subsystem with CALCulate:RESistivity subsystem) 2. To control the deviation measurement mode (CALCulate1:MATH subsystem) 3. To control the comparator function (CALCulate1:LIMit subsystem) 4. To control the current monitor function (CALCulate2 subsystem) 5.
4339B CALCulate Subsystem :CALCulate1:FORMat f REAL j SRESistivity j VRESistivity g Speci es or queries the post-processing of measurement data, which is measured using the speci ed by the [SENSe]:FUNCtion command.
CALCulate Subsystem 4339B :CALCulate1:LIMit:FAIL? Returns the comparison result. (Query only) Query response is 1 or 0. Where, 1 Comparison result is FAIL. 0 Comparison result is PASS. :CALCulate1:LIMit:LOWer[:DATA] Sets or queries lower limit data. is, numeric Real. Range is 09.921037 to 9.921037 . Query response is a numeric value in format. :CALCulate1:LIMit:LOWer:STATe f ON j OFF j 1 j 0 g Sets or queries if the lower test limit is enabled.
4339B CALCulate Subsystem :CALCulate1:MATH:EXPRession:CATalog? Returns available parameters which are used with CALCulate1:MATH:EXPRession:NAME command. (query only) For more information, see CALCulate1:MATH:EXPRession:NAME command. Query response is {DEV|PCNT}. :CALCulate1:MATH:EXPRession:NAME f DEV j PCNT g De nes or queries the expression used for deviation measurement, which is enabled by CALCulate1:MATH:STATe. The reference value can be de ned using the DATA[:DATA] command.
CALCulate Subsystem :CALCulate1:RESistivity:GLENgth Speci es or queries the distance between the main electrode and guard electrode (ring) for the resistivity measurement. is, numeric 0.00001 to 0.9999 (m) Query response is a numeric value in format. :CALCulate1:RESistivity:STHickness Speci es or queries the thickness of the sample for the resistivity measurement. is, numeric 0.00001 to 0.
4339B CALCulate Subsystem :CALCulate3:BEEPer fONjOFFj1j0g Sets or queries if the beeper of the measurement sequence program is enabled. ON, 1 Two beeps are emitted when the sequence program is completed. OFF, 0 No beep is emitted. Query response is 1 or 0. Note "1" of CALCulate1 can be omitted. "2", "3" of CALCulate2, CALCulate3 cannot be omitted. The short must be CALC2, CALC3.
4339B CALibration Subsystem COMMAND CALibration [:ALL]? :AUTO PARAMETER :CALibration[:ALL]? Performs a calibration, and then returns an error code. The error code is referred to \Messages" at the back of this manual (An error code of \0" indicates no failure). :CALibration:AUTO fONjOFFj1j0g Sets or queries if the o set-error canceling function is enabled.
4339B DATA Subsystem DATA Subsystem The DATA subsystem is used to store data to the 4339B's data bu er and to read data in the data bu er. COMMAND DATA [:DATA] [:DATA]? :FEED :CONTrol :POINts :DATA[:DATA] REF, DBUF, IMON, TMON DBUF, DBUF,fALWaysjNEVerg DBUF, value> Enters or queries the reference value for the deviation measurement, which is controlled by :CALCulate:MATH subsystem commands. is, numeric 09.
DATA Subsystem 4339B :DATA[:DATA]? IMON Returns the current monitor value. This query needs parameter IMON which is the name of the data bu er to read. Query response is a numeric value in format. :DATA[:DATA]? TMON Returns the timer data of the measurement sequence program (in seconds, or in %) when the time display function is ON. This query needs parameter TMON which is the name of the data bu er to read. Query response is a numeric value in format.
4339B DISPlay Subsystem DISPlay Subsystem COMMAND DISPlay [:WINDow] [:STATe] :TEXT1 :DIGit :PAGE :PREFix :TEXT2 :PAGE PARAMETER {3|4|5} {1|2|3} {1|2|3|4|5|6} :DISPlay[:WINDow][:STATe] fONjOFFj1j0g Sets the display ON or OFF, or queries whether the display is set to ON or OFF. Where, ON or 1 Sets the display ON. OFF or 0 Sets the display OFF (blank). Query response is 1 or 0. :DISPlay[:WINDow]:TEXT1:DIGit f3j4j5g Sets or queries the display digits of the measurement data.
DISPlay Subsystem :DISPlay[:WINDow]:TEXT2:PAGE f1j2j3j4j5j6g Sets or queries the Measurement Settings display mode. 1 Test voltage and Current limit 2 Averaging rate and Trigger delay time 3 Sequence mode and charge time 4 Sequence interval time and Memory size 5 Comparator limits 6 Test voltage and Current monitor value Query response is 1,2,3,4,5, or 6. Note "1" of :TEXT1 can be omitted, but "2" of :TEXT2 cannot be omitted.
4339B FETCh? Query FETCh? Query :FETCh? Retrieves a measurement value taken using the INITiate subsystem commands and places the measurement data into the 4339B's output bu er.
4339B FORMat Subsystem COMMAND FORMat [:DATA] PARAMETER fASCiijREAL[,64]g :FORMat[:DATA] f ASCii j REAL[,64] g Sets the data format for transferring numeric and array information. ASCii Sets the data transfer format to ASCII. REAL[,64] Sets the data transfer format to IEEE 754 oating point numbers of the speci ed length of 64-bit. For details on data transfer formats, see \Data Transfer Format". Query response is ASC or REAL,64.
4339B INITiate Subsystem INITiate Subsystem The INITiate subsystem controls the initiation of the trigger system. All trigger sequences are initiated as a group. The detailed description of the trigger system is given in \Trigger System". COMMAND INITiate :CONTinuous [:IMMediate] PARAMETER :INITiate[:IMMediate] Causes all sequences to exit Idle state and enter Initiate state. Then the trigger system initiates and completes one full trigger cycle.
4339B OUTPut Subsystem The OUTPut subsystem controls the voltage source. COMMAND OUTPut [:STATe] PARAMETER :OUTPut[:STATe] f ON j OFF j 1 j 0 g Controls or queries whether the source voltage is ON or OFF. ON or 1 Turns the source voltage ON. OFF or 0 Turns the source voltage OFF. Query response is 1 or 0. Warning Sending :OUTPut[:STATe] { ON | 1 } may cause the 4339B to output dangerous voltage levels up to 1000 Vdc.
4339B SENSe Subsystem SENSe Subsystem COMMAND [SENSe] :AVERage :COUNt [:STATe] :CORRection :COLLect [:ACQuire] :DATA? [:STATe] :CURRent :APERture :RANGe :AUTO [:UPPer] :FUNCtion :CONTact :DATA? :LIMit? :OFFSet :VERify PARAMETER OFFSet OFFSet, SCAPacitance [MSjS] [PAjNAjUAjA] [PFjF] [:SENSe]:AVERage:COUNt Sets or queries the averaging rate.
SENSe Subsystem 4339B [:SENSe]:CONTact:DATA? Returns the capacitance data at dut mesurement. The contact check function uses this capacitance data. Refer to \Reading the Limit Data of Contact Check Function" in Chapter 2 for detail. (query only) Query response is a numeric value in format. The data is incorrect if the measurement range was changed after dut measurement. Note The accuracy of the capacitance data is not guaranteed. [:SENSe]:CONTact:LIMit? Returns the contact check limit data.
4339B SENSe Subsystem [:SENSe]:CORRection:DATA? f OFFSet j SCAPacitance g Returns the OPEN correction data (DC current measurement value) or capacitance measurement data obtained by performing the OPEN correction. The capacitance measurement data is used for the contact check function.(Query only) OFFSet Returns the open correction data. SCAPacitance Returns the capacitance measurement data. Query response is a numeric value in format.
SENSe Subsystem [:SENSe]:FUNCtion Sets or queries the measurement parameter. When an *RST command is executed, this is set to "RESistance[:DC]". is 'CURRent[:DC]' Current measurement 'RESistance[:DC]' Resistance measurement To measure surface or volume resistivity, calculate the value using the :CLACulate:FORMat command. Query response is "CURR" or "RES".
4339B SOURce Subsystem SOURce Subsystem COMMAND SOURce :CURRent :LIMit [:AMPLitude] :VOLTage [:LEVel] [:IMMediate] [:AMPLitude] PARAMETER [MA|A] [V|KV] :SOURce:CURRent:LIMit[:AMPLitude] [MAjA] Sets or queries the current limit value. is, numeric 0.5 mA(default), 1 mA, 2 mA(at test voltage 0 to 500 V only), 5 mA(at test voltage 0 to 250 V only), 10 mA(at test voltage 0 to 100 V only) Query response is a numeric value, in format.
4339B STATus Subsystem The STATus subsystem commands control the Operation Status and Questionable Status registers in the status-reporting structures (See \Status Reporting Structure"). COMMAND STATus :OPERation :CONDition? :ENABle [:EVENt]? :PRESet :QUEStionable :CONDition? :ENABle [:EVENt]? PARAMETER :STATus:OPERation:CONDition? Returns the contents of the condition register of the Operation Status Register group. Reading the condition register does not clear it.
4339B :STATus:QUEStionable:ENABle Sets or queries the enable register of the Questionable Status Register group. is, numeric Decimal expression of the contents of the register. The 4339B has no operation to report the event to the Questionable Status Condition Register. Query response is a numeric value in format. :STATus:QUEStionable[:EVENt]? Returns the contents of the event register of the Questionable Status Register group.
4339B SYSTem Subsystem The SYSTem subsystem reports the SCPI version and error, sets the beeper, disables front-panel key input, and sets the line frequency. COMMAND SYSTem :BEEPer [:IMMediate] :STATe :ERRor? :KLOCk :LFRequency :PRESet :VERSion? PARAMETER :SYSTem:BEEPer[:IMMediate] Produces a beep immediately. (no query) :SYSTem:BEEPer:STATe f ON j OFF j 1 j 0 g Sets or queries if the beeper is enabled.
4339B SYSTem Subsystem :SYSTem:PRESet Resets the 4339B to its default states. (No query) The default states are given in Table 3-3. :SYSTem:VERSion? Returns the value corresponding to SCPI version to which the instrument complies. (Query only) Query response is YYYY.V.
4339B TRIGger Subsystem The TRIGger subsystem controls the measurement trigger function. For a detailed description about the trigger sequence controlled by the TRIGger subsystem, see \TRIG Event Detection State" in \Trigger System", later in this chapter.
4339B TRIGger Subsystem :TRIGger[:SEQuence1]:SOURce fBUSjEXTernaljINTernaljMANualjTIMerg Sets or queries a trigger source. Only one source can be speci ed at a time. The TRIGger[:SEQuence1]:SOURce command works with the ARM[:SEQuence1][:LAYer]:SOURce command and controls the measurement sequence mode. The selections of the TRIGger:SOURce and ARM:SOURce is shown in Table 5-1. ARM:SOUR1 IMMediate Table 5-1.
TRIGger Subsystem 4339B :TRIGger[:SEQuence1]:TIMer [MSjS] Sets the period of the internal periodic signal source. is, numeric 10 ms to 999 s The range and resolution are as follows: Range 10 ms Interval Time < 10 s 10 s Interval Time < 100 s 100 s Interval Time 999 s Resolution 1 ms 10 ms 100 ms Query response is a numerical value in format. When the *RST command is executes, period is sets as follows.
3OPC 4339B Common Commands 3CLS Clears the the Operation Status Event Register, and the Standard Event Status Register. (No query) 3ESE Sets or queries the bits of the Standard Event Status Enable Register. is, numeric A decimal expression of the contents of the Standard Event Status Enable Register Query response is a numeric value in format. 3ESE? Queries the contents of the Standard Event Status Enable Register.
3OPC 4339B 3RCL Recalls the instrument state previously stored in the speci ed register number. The 4339B has 10 available storage registers. (No query) is, numeric 0 to 9 3RST Returns the 4339B to its default state and sets the following commands. INITiate:CONTinuous OFF [SENSe:]CORRection[:STATe] OFF SYSTem:KLOCk OFF (No query) 3SAV Saves the instrument state in the speci ed register number. The 4339B has 10 available storage registers.
4339B Common Commands 3TST? Executes an internal self-test and returns the test result as the sum of the error codes of all existing errors. If there is no error the 4339B returns a 0. 1. 2. 3. 4. 5. 6. Test Item RAM EPROM Calibration data (EEPROM) User's data (EEPROM) AD converter Backup RAM Error Code 1 2 4 8 16 32 Query response is a numeric value in format. 3WAI Makes the 4339B wait until all previously sent commands are completed.
Status Reporting Structure 4339B Status Reporting Structure This section provides information about the status reporting structure for Service Request functions which correspond to SCPI. Figure 5-2. Status Reporting Structure Service Request (SRQ) The 4339B can send an SRQ (Service Request) control signal when it requires the controller to perform a task. When the 4339B generates an SRQ, it also sets Bit 6 of the Status Byte Register, RQS (Request Service) bit.
4339B Status Reporting Structure Status Byte Register The Status Byte Register is composed of eight bits that summarize an overlaying status data structure. The Status Byte Register can be read using either 3STB? or SPOLL, which return a decimal expression of the contents of the register (equal to the sum of the total bit weights of all the bits set to \1"). Figure 5-3. Status byte Register Bit No. Bit Weight 7 6 128 64 5 4 32 16 3 2-0 8 Table 5-2.
Status Reporting Structure 4339B Standard Event Status Register The Standard Event Status Register is frequently used and is one of the simplest. You can program it using GPIB common commands, *ESE and *ESR?. Refer to *ESE command and *ESR? command in \Command Reference". Figure 5-4. Standard Event Status Register Table 5-3. Standard Event Status Register Assignments Bit No.
4339B Status Reporting Structure Standard Operation Status Group The 4339B provides two Standard Operation Status groups | Operation Status Register group and Questionable Status Register group | which can be accessed using the STATus subsystem commands. (Refer to STATus subsystem in \Command Reference".) This group is used in advanced programming. The individual bit assignment of these registers are given in \Operation Status Register" and \Questionable Status Register" later in this section.
Status Reporting Structure 4339B Operation Status Register The Operational Status Register group provides information about the state of the 4339B measurement system. Table 5-4. Operation Status Condition Register Assignments Description Bit No. Bit Weight 9-15 8 7 256 128 6 64 5 32 4 3 2 16 1 2 0 1 4 Always 0 DBUF is full | this bit is set when DBUF is full. Correcting | This bit is set when the correction data measurement is in progress.
4339B Trigger System Trigger System This section provides information about the trigger system of the 4339B. SCPI de nes a common trigger model for several types of instruments. The trigger system allows you to have speci c control of your measurements. Information on the trigger system requires more technical expertise than most other topics covered in this chapter. But you can avoid having to learn the information in this section by using the :INITiate commands to make your measurements.
Trigger System 4339B Idle State The trigger system remains in the idle state until it is initiated by :INITiate[:IMMediate] or :INITiate:CONTinuous ON commands. Once one of these conditions is satis ed, the trigger system exits downward to the initiate state. Note that *RST sets INITiate:CONTinuous OFF. Initiate State If the trigger system is on a downward path, it travels directly through the initiate state without restrictions.
4339B TRIG Event Detection State Trigger System Figure 5-8. Inside a TRIG Event Detection State SOURce DELay IMMediate COUNt The :TRIGger[:SEQuence1]:SOURce command speci es which particular input can generate the event required to continue the downward path. The :TRIGger[:SEQuence1]:DELay command speci es a time duration between the recognition of an event and the downward exit. Sending an *RST command sets DELay to zero.
Data Transfer Format 4339B Data Transfer Format ASCii Numeric data is transferred as ASCii bytes in (integer), ( xed point) or ( oating point) formats, as appropriate. The numbers are separated by commas as speci ed in IEEE 488.2. Figure 5-9. NR1 Format Figure 5-10. NR2 Format Figure 5-11. NR3 Format For example, expressions for \1000" are, +1000 format +1000.0 format +1.
4339B Data Transfer Format REAL Figure 5-12. Real Data Format This data eld is initiated by a unique code, the number sign (#). The second byte, \" designates the number of the bytes for the \". \" designates \" length. \&^END " is the response message terminator. \" contains the data in IEEE 754 oating point numbers of 64-bits. This is the same data format used by the HP Technical computers, such as the HP 9000 series 200/300 computers.
Command Summary 4339B Command Summary Command 3CLS 3ESE 3ESR? 3IDN? 3LRN? 3OPC 3RCL ABORt ARM Table 5-6. Common Commands Parameter [query only] [query only] [query only] [no query] Command 3RST 3SAV 3SRE 3STB? 3TRG 3TST? 3WAI Parameter Note [no query] [no query] [query only] [no query] [query only] [no query] Table 5-7.
4339B Command Summary Table 5-7.
Command Summary 4339B Table 5-7.
6 Application Measurement Introduction This chapter provides actual measurement examples for various types of devices.
Measuring Insulation Resistance of Capacitors 4339B Measuring Insulation Resistance of Capacitors This example shows insulation resistance measurement of a chip capacitor after it has been charged for 1 minute using the 16339A Component Test Fixture with the SMD module. Warning Do NOT touch the UNKNOWN terminals or the electrodes of the accessory, when the High Voltage indicator is ON, the 4339B outputs dangerous voltage levels up to 1000 Vdc.
4339B Measuring Insulation Resistance of Capacitors 6. Perform the OPEN correction. a. Remove the DUT if connected, and set the SMD module's electrode contact spacing the same as the DUT's length, tightening the screw to hold this interval between the electrodes. b. Close the top cover. c. Select the Current measurement mode. Press to display the measurement parameter selection menu. Select I using or and press . d. Apply the source voltage by pressing . e.
Measuring Insulation Resistance of Capacitors 9. Set the measurement sequence con guration. Press . Select Chrg using or for example 60 (s). Press twice. 4339B and press 10. Select the single mode as the measurement sequence mode. Press . Select Single using or and press . Enter the charge time, . 11. Start the measurement sequence. . The Seq Running annunciator( 9 ) is turned ON, and the 4339B applies the test Press voltage across the capacitor for 60 seconds.
4339B Measuring Resistivity of Insulation Materials Measuring Resistivity of Insulation Materials This section provides measurement example of resistivity of insulation lm using the 16008B Resistivity Cell with the 50 main electrode. Warning Do NOT touch the UNKNOWN terminals or the electrodes of the accessory, when the High Voltage indicator is ON, the 4339B outputs dangerous voltage levels up to 1000 Vdc.
Measuring Resistivity of Insulation Materials 4. Perform calibration. Press . Select ExecCal using 4339B or and press . 5. Perform the OPEN correction. a. Turn the load knob counterclockwise until the upper electrode does not move. b. Close the top cover. c. Select the Current measurement mode. Press to display the measurement parameter selection menu. Select I using or and press . d. Apply the source voltage by pressing . e. Wait until the current has stabled to within 0.5 pA.
4339B Measuring Resistivity of Insulation Materials 7. Turn the load knob to apply pressure by the upper and lower electrodes on the DUT, and then close the top cover. Turning load knob applies load pressure on the DUT of up to approximately 10 kg. If you do not need to apply a speci c pressure, DUT must at least rmly held by the upper and lower electrodes. 8. Set the resistivity cell parameter a. Press . b. Enter DUT thickness. Select Thickenss using or example 0.13 (mm), and press . c.
Measuring Resistivity of Insulation Materials 4339B 11. Measuring the volume resistivity. a. Turn the Volume/Surface selector of the 16008B to \Volume". b. Press . Select Rv using or and press . c. Press . This starts the volume resistivity measurement sequence. Measured volume resistivity value will be displayed after 60 seconds charge time. Note If you want to abort the measurement sequence, press . 12. Measuring surface resistivity. a. Turn the Volume/Surface selector to \Surface".
4339B Measuring Insulation Resistance Time Characteristics Measuring Insulation Resistance Time Characteristics of Electro-Mechanical Components This section provides an example of measuring electro-mechanical components like switches, connectors, or relays. This example measures insulation resistance (IR) time characteristics of opened switch contacts using the continuous measurement sequence mode, to obtain time characteristics.
Measuring Insulation Resistance Time Characteristics 5. Perform the calibration. Press . Select ExecCal using or 4339B and press . 6. Perform the OPEN correction. a. Remove the alligator clips and banana cables from the INPUT terminal. b. Close the top cover. c. Select the current measurement mode. Pressing . Select I using or and press . d. Apply test voltage by pressing . e. Wait until the current has stabled to within 0.5 pA. (Displayed I value do not change by more than 0.5 pA.) f.
4339B Measuring Insulation Resistance Time Characteristics 9. Set sequence mode parameters. Press . a. Set the charge time. Select Chrg. Enter the charge time value, in this example 60 (s). Press . b. Set the interval time and memory size. Select Intvl. Enter the interval time value, in this example 10 (s). Press . Then enter the number of measurements (equivalent to the memory size), in this example 60. Press . c. Select the continuous mode as the measurement sequence mode. Press .
7 Measurement Basics Introduction This chapter provides information that will help you to work with your 4339B more e ectively.
Insulation Resistance Measurement 4339B Insulation Resistance Measurement Residual Charge E ect Once a high resistance insulation material is charged with a voltage, a residual voltage will still remain on the material for a while after the charging voltage is removed.
4339B Resistivity Resistivity Volume Resistivity In general, volume resistivity is de ned as the resistance per unit volume.
High Capacitance DUT Measurement 4339B Surface Resistivity Surface Resistivity is de ned as resistance per unit surface area. Similar to volume resistivity, surface resistivity is the resistance of a square of a material whose side length is 1 m, ( one square meter).
4339B High Capacitance DUT Measurement High Capacitance DUT Measurement When the 4339B measures a DUT that has high capacitance, AC noise creates major instability factors in the measurement because capacitance is sensitive to AC noise. If you want to measure the resistance of a high-capacitance DUT maintaining more than 10 % of S/N (Signal/Noise) ratio, the maximum capacitance of the DUT's for each measurement range are listed in Table 7-1. Table 7-1.
High Capacitance DUT Measurement 4339B In addition, Table 7-3 shows the appropriate resistor for each range and measurement time mode. Measurement Range Table 7-3.
4339B High Capacitance DUT Measurement Figure 7-3 shows the measurement con guration for this application. Figure 7-3. Capacitance Measurement Con guration Where, Rs Output Resistance of the 4339B Rx Parallel resistance of the DUT Cx Parallel capacitance of the DUT Ri Input resistance of the 4339B Ro Additional Series Output Resistance of the 16339A Necessary charging time is approximately led by following equation.
8 Speci cations This chapter contains a list of speci cations for reference and performance veri cations. When shipped from the factory, the 4339B meets the speci cations listed in this chapter. This chapter also includes supplemental characteristics. These characteristics are not speci cations but are typical characteristics included as supplemental information for the operator.
Speci cations Speci cations 4339B These speci cations are the performance standards or limits against which the instrument is tested. When shipped from the factory, the 4339B meets the speci cations listed in this section. The speci cation test procedures are covered in Chapter 9. Measurement Parameters R (dc resistance), I (dc current), s (surface resistivity), v (volume resistivity) Measurement Conditions DC Test Voltage (Vs) Range 0 to 1000 V, 0.1 V steps for Vs 200 V or 1.
4339B Trigger Mode Speci cations Internal, Manual, External, and Bus Trigger Delay Time 0 to 9.999 seconds in 0.001 seconds steps Measurement Range R: 12103 to 1.
Speci cations 4339B Measurement Accuracy The following conditions must be met for measurement accuracy to apply: 1. Warm up time: 30 min 2. Ambient temperature: 23 6 5 C 3. O set-error canceling is ON. 4. OPEN correction has been performed. 5. Self calibration has been performed. Table 8-1.
4339B Speci cations Table 8-1.
Speci cations 4339B Measurement Support Functions Display 40-digit 2 2-lines LCD display. Capable of displaying: measurement values, comparator/contact check decisions, comparator limits, control settings, self test messages, and annunciators Correction OPEN correction Eliminates measurement errors due to stray capacitance in the test cable and residual resistance in the test xture. Self calibration: Calibrates the instrument o set errors due to the change of ambient temperature.
4339B Continuous Memory Capability Speci cations If the instrument is turned OFF, or if a power failure occurs, the instrument settings are automatically saved. Key Lock Disables key input from the front panel. General Power Requirements 100/120/220/240 V 610%, 47 to 66 Hz 45 VA max Operating Temperature, Humidity, and Altitude 0 to 45 C, 95% RH @ 40 C, 0 to 2000 meters Condensation must be avoided.
Supplemental Performance Characteristics 4339B Supplemental Performance Characteristics The supplemental performance characteristics are listed below. These characteristics are not speci cations but are typical characteristics included as additional information for the operator.
4339B Supplemental Performance Characteristics DC Test Voltage Settling Slew rate until the output voltage reaches 90 % of the set voltage (Set V) is, V'/t' = 10 V/ms Where V' and t' is shown in the following gure.
9 Maintenance Introduction This chapter provides information on how to maintain the 4339B. 4339B maintenance consists of performance tests and functional tests. Test Equipment Table 9-1 lists the recommended equipment for performing 4339B maintenance. Equipment DC Voltmeter RC Box RC Box Adapter V Measurement Adapter Cable Cable Handler Interface Tester Table 9-1. Required Equipment Requirements Recommended Model Qty. Use1 Voltage Range: 0 V to 1000 V Accuracy: < 0.
Performance Tests 4339B Performance Tests Introduction This section provides the test procedures used to verify that the 4339B's speci cations listed in Chapter 8, General Information, of this manual are met. All tests can be performed without access to the interior of the instrument. The performance tests can also be used to perform incoming inspection, and to verify that the 4339B meets its performance speci cations after troubleshooting or adjustment.
4339B Performance Tests Performance Test Record Record the performance test results in the test record at the end of this chapter (Photocopy the test record and use the photocopy). The test record lists all test speci cations, their acceptable limits, and measurement uncertainties for the recommended test equipment. Test results recorded during incoming inspection can be used for comparison purposes during periodic maintenance, troubleshooting, and after repair or adjustment.
Performance Tests 4339B Source Voltage Accuracy Test The 4339B's source voltage is measured with a DC voltmeter. Speci cation Source Voltage (Vs) Accuracy: 6 (0.16 % + 100 mV) (Vs 200 V) 6 (0.16 % + 500 mV) (Vs > 200 V) Test Equipment Description Recommended Model Multimeter V Measurement Adapter Banana-Banana Jumper 3458A PN 04339-65006 11508A Procedure Warning When the High Voltage indicator is lit, the 4339B outputs high voltages of up to 1000 Vdc maximum. 1.
4339B Performance Tests 3. Press the 3458A Multimeter's 4DCV5 to set the measurement mode to DC voltage. 4. Set the V Measurement Adapter's ID switches and Output switch as follows: Switch Setting ID1 0 ID2 0 ID3 0 ID4 1 Output Enable 5. On the 4339B, press to turn the source voltage ON. (V Output indicator turns ON.) 6. Record the multimeter reading on the calculation sheet. (Initial output voltage setting is 0 V.) 7. Press to turn the source voltage OFF. (V Output indicator turns OFF.) 8.
Performance Tests 4339B Ammeter O set Voltage and Input Resistance Test The 4339B's ammeter o set voltage and input resistance are measured. Speci cation 0.5 mV 1 k 6 5% Ammeter O set Voltage: Ammeter Input Resistance: Test Equipment Description Recommended Model Multimeter RC Box Adapter Dual Banana-BNC(m) Cable 3458A PN 04339-65005 PN 11001-60001 Procedure Ammeter O set Voltage Test. 1. Reset the 4339B using the following procedure: a. Press to display the system reset menu. b.
4339B Performance Tests 5. Press the 4339B's to set the measurement time to SHORT. The current measurement time setting is indicated by the Meas Time annunciator( 9 ). 6. Set the measurement range to 100 A using the following procedure: a. Press to display the measurement range setup menu. b. Press 7. Press until 100 A appears, and press . to set the trigger mode to Manual. The current trigger mode setting is indicated by the Trigger annunciator( 9 ). 8.
Performance Tests 4339B Current Measurement Accuracy Test The 4339B measures the current through the calibrated RC Box, and the measured values are compared with the current values calculated from the RC box's calibration value and the voltage across the standard. Speci cation 6 0.4 % (See Chapter 8 General Information for Basic Measurement Accuracy: details.
4339B Performance Tests Figure 9-4. Current Measurement Accuracy Test Setup 5. Set the RC Box Adapter switch to the F position. 6. Press the 3458A Multimeter's 4DCV5 to set the measurement mode to DC voltage. 7. Press . Select ExecCal and press to perform the calibration. 8. Set the source voltage to 10 V using the following procedure: a. Press to display the source voltage setup menu. b. Press to set the source voltage to 10 V. c. Con rm that +10 V is displayed on the right side of the LCD. 9.
Performance Tests 4339B Figure 9-5. RC Box Connection 13. Press . Select I and press 14. Press to set the measurement time to LONG. The current measurement time setting is to set the measurement parameter to current (I). indicated by the Meas Time annunciator( 9 ). 15. Press to set the trigger mode to Manual. The current trigger mode setting is indicated by the Trigger annunciator( 9 ). 16. Set the RC Box resistor to 1010 . 17. Set the source voltage to 1 V using the following procedure: a.
4339B Performance Tests Table 9-3.
Performance Tests 4339B Resistance Measurement Accuracy Test The 4339B measures the resistance values of the calibrated RC Box, and these measurements are compared with the RC Box's calibration values. Speci cation 6 0.6 % (See Chapter 8 General Information for Basic Measurement Accuracy: details.
4339B Performance Tests Figure 9-6. Resistance Measurement Accuracy Test Setup (Floating) 5. Set the RC Box Adapter switch to the F position. 6. Press . Select ExecCal and press to perform the calibration. 7. Set the source voltage to 100 V using the following procedure: a. Press to display the source voltage setup menu. b. Press to set the source voltage to 100 V. c. Con rm that + 100 V is displayed on the right side of the LCD. 8. Press to turn the source voltage ON. (V Output indicator turns ON.) 9.
Performance Tests 4339B Figure 9-7. RC Box Connection (Floating) 12. Press to set the measurement time to SHORT. The current measurement time setting is indicated by the Meas Time annunciator( 9 ). 13. Press to set the trigger mode to Manual. The current trigger mode setting is indicated by the Trigger annunciator( 9 ). 14. Set the RC Box resistor to 106 . 15. Press to turn the source voltage ON. (V Output indicator lights.) 16. Press to measure. 17. Record the 4339B reading on the calculation sheet.
4339B Performance Tests Table 9-4. Resistance Measurement Accuracy Test Settings Resistor Voltage Measurement Setting Setting Time 106 100 V Short 107 100 V Long 108 100 V Long 109 100 V Long 1010 100 V Long 1011 100 V Long 11 10 100 V Short 1011 10 V Long Resistance Measurement Accuracy Test (Grounded Device). 21. Set up the equipment as shown in Figure 9-9. Figure 9-9. Resistance Measurement Accuracy Test Setup (Grounded) 22. Set the RC Box Adapter switch to the G position. 23.
Performance Tests 24. Press to turn the source voltage ON. (V Output indicator turns ON.) 25. Press 26. Press 4339B . Select OpenMeas and press to perform the OPEN correction. to turn the source voltage OFF. (V Output indicator turns OFF.) 27. Connect the RC Box to the RC Box Adapter as shown in Figure 9-10. Figure 9-10. RC Box Connection (Grounded) 28. Set the RC Box resistor to 107 . 29. Press to turn the source voltage ON. (V Output indicator turns ON.) 30. Press to measure. 31.
4339B Performance Tests Calculation Sheet Source Voltage Accuracy Test Source Voltage Setting Multimeter Reading [a] Test Result Equation 0V 10 V 25 V 50 V 100 V 200 V 201 V 250 V 500 V 1000 V V V V V V V V V V V a a 0 10.000 V a 0 25.000 V a 0 50.000 V a 0 100.00 V a 0 200.00 V a 0 201.00 V a 0 250.00 V a 0 500.00 V a 0 1000.
Performance Tests 4339B Current Measurement Accuracy Test Test Current Resistor Voltage Setting Setting 100 pA 1 nA 10 nA 100 nA 1 A 10 A 100 A 1010 109 108 107 106 105 105 Multimeter Reading [a] 1V 1V 1V 1V 1V 1V 10 V 4339B Reading [b] V V V V V V V Test Result Equation1 pA b 0 a 4 cv2 nA b 0 a 4 cv3 nA b 0 a 4 cv4 nA b 0 a 4 cv5 A b 0 a 4 (cv6+1 k ) A b 0 a 4 (cv7+1 k ) A b 0 a 4 (cv7+1 k ) 1 1 k in the test result equation is the input resistance of the 4339B's ammeter.
4339B Performance Tests Performance Test Record Agilent Technologies 4339B High Resistance Meter Serial No.: Date: Temperature: Humidity: Tested by: Source Voltage Accuracy Test Source Voltage Test Limits Setting 60.10 V 60.12 V 60.14 V 60.18 V 60.26 V 60.42 V 60.82 V 60.90 V 61.3 V 62.1 V 0V 10 V 25 V 50 V 100 V 200 V 201 V 250 V 500 V 1000 V 1 Test Result = Measured Value Measurement Uncertainty Test Result1 V V V V V V V V V V 6 0.000002 V 6 0.00018 V 6 0.00073 V 6 0.00073 V 6 0.0022 V 6 0.
Performance Tests 4339B Current Measurement Accuracy Test Test Current Test Limits Measurement Time 100 pA 1 nA 10 nA 100 nA 1 A 10 A 100 A 62.6 pA 60.0094 nA 60.063 nA 60.53 nA 60.0043 A 60.036 A 60.60 A 1 Test Result = 4339B Reading Test Result1 Measurement Uncertainty pA nA nA nA A A A Long Long Long Long Long Long Short 60.56 pA 60.0016 nA 60.011 nA 60.10 nA 60.0005 A 60.007 A 60.
4339B Functional Test Functional Test Introduction This section provides the test procedures used to verify that the 4339B performs its designed functions. The functional tests can be used for post-repair function veri cation. Test Equipment Table 9-1 lists the test equipment required to perform the tests described in this section. Equipment which equals or surpasses the key required speci cations of the recommended equipment may be used.
Functional Test 4339B Handler Interface Functional Test The 4339B's handler interface function is tested using the built-in selftest and the handler interface tester. Test Equipment Description Recommended Model Handler Interface Tester 04339-65007 Procedure Initial Setup. 1. Turn the 4339B OFF. 2. Set the LED PW switch on the Handler Interface Tester to OFF, and set the IN1, IN2, IN3, and IN4 switches to 0. 3. Set up the equipment as shown in Figure 9-11 Figure 9-11.
4339B Functional Test 11. Press the IN5 switch on the Handler Interface Tester, and con rm that the 4339B is triggered. Handler Interface Output Test. 12. Start the handler interface output test using the following procedure: a. Press to display the con guration setting menu. b. Select more using c. Select Svc using d. Select HNDL using or or or and press and press and press . . . 13. Set the LED PW switch on the Handler Interface Tester to ON.
Functional Test 4339B Contact Check Functional Test The 4339B's contact check function is con rmed. Test Equipment Description Recommended Model RC Box RC Box Adapter 16340A 04339-65005 Procedure 1. Reset the 4339B using the following procedure: a. Press to display the system reset menu. b. Select Yes using or and press . 2. Set up the equipment as shown in Figure 9-13. Figure 9-13. Contact Check Functional Test Setup 3. Set the RC Box Adapter switch to the G position. 4.
4339B 6. Press Functional Test and select ON/OFF to turn the contact check function ON. 7. Con rm that the 4339B displays N.C. (No Contact). 8. Set the 16340A RC Box resistor to 104 . 9. Connect the RC Box to the RC Box Adapter as shown in Figure 9-14. Figure 9-14. Contact Check Functional Test Setup 10. Con rm that the N.C. disappears and the 4339B displays the resistance measurement value.
A Manual Changes Introduction This appendix contains the information required to adapt this manual to earlier versions or con gurations of the 4339B than the current printing date of this manual. The information in this manual applies directly to the 4339B serial number pre x listed on the title page of this manual. Manual Changes To adapt this manual to your 4339B, see Table A-1 and Table A-2, and make all the manual changes listed opposite your instrument's serial number and rmware version.
4339B Serial Number Agilent Technologies uses a two-part, ten-character serial number that is stamped on the serial number plate (see Figure A-1) attached to the rear panel. The rst ve characters are the serial pre x and the last ve digits are the su x. Figure A-1.
4339B Change 1 Delete the following function. Limit Value Reading function ( LimitC) O set Setting function ( OfsEnt) Dut Capacitance Reading function ( MeasC) Delete the following GPIB command. :SENSe:CONTact:DATA? :SENSe:CONTact:LIMit? :SENSe:CONTact:OFFSet Change 2 Change Table C-1 and Table C-2 to Table A-3 and Table A-4 respectively. Table A-3. OVLD/Over-Current/N.C. Operations (Meas.
4339B Table A-4. OVLD/Over-Current/N.C. Operations(Meas. parameter: I) Display Data Comprtr Handler Output GPIB Output1 Mode Mode OVLD (Overload) OVLD HIGH Over-Current OV CURR HIGH N.C. N.C. (No-Contact) OVLD & N.C. N.C. OVLD & OV CURR Over-Current N.C. N.C. HIGH 2 : 1 /HI : 9.9E37 3 : 2 /HI 2 : 4 & : 9.9E37 /NOT READY 3 : 2 2 : 2 /NO CONTACT : 9.9E37 3 : 8 2 : 3 /HI : 9.9E37 3 : 2 /HI 2 : 5 & : 9.
B Handler Interface Installation This appendix provides the electrical characteristics of the handler interface signals, and how to set up the handler interface board. Electrical Characteristics Output Signals Each DC output is isolated using open collector output opto-isolators. The output voltage of each line is enabled by putting pull-up resistors on the main board, and by connecting the pull-up resistors to an externally applied DC voltage.
4339B Figure B-1.
4339B Figure B-2.
4339B Input Signals The DC isolated input signals are connected to the cathodes of the LEDs in the opto-isolators. The anodes of the LED's are powered by an external voltage source (EXT DCV2). The OFF state voltage (high level) of the DC isolated input signals depends on the pull-up voltage powered by an external; voltage source (EXT DCV2). (The input current is restricted by using a switch on the main board.) The electrical characteristics of the input signals are listed in Table B-2.
4339B Procedure Setting Up the Handler Interface Board Caution SUSCEPTIBLE TO DAMAGE FROM ESD. Perform the following procedures only at a static-safe workstation and wear a grounding strap. Caution Be careful not to short the circuit when performing the following procedures. For example: Solder cleanly and carefully. Guard against hair or dust getting on the circuit. Do not damage the boards, wires, or parts on the board.
Procedure 4339B Figure B-4. Top View of 4339B with Cover Removed 4. Remove the A2 CPU bard assembly. a. Disconnect the four atcable assemblies from the A2 board. b. Remove the four screws that secure the A1 board to the stud. c. Remove the A2 board form the chassis.
4339B Procedure 5. Remove the A1 board assembly. a. Disconnect the following cable assemblies from the A1 board. i. Four cable assemblies form the front panel ii. The cable assembly from the transformer iii. Two cable assemblies form the DC-DC Converter b. Disconnect the four optical ber cables which are connected to the A# Ammeter Board Assembly c.
Procedure 4339B Figure B-5.
4339B Procedure Table B-3. Pull-up Resistor Location Socket No. Signal Name Signal Type J11-1 /INDEX Control Signal (5 to 15 V) J11-2 /EOM J11-3 /ALARM J11-4 /NOT READY J11-5 /NO CONTACT Comparison Signal (5 to 24 V) J11-6 /LO J11-7 /HI J11-8 /IN J11-9 Not Used J11-10 J11-11 6. Mount the pull-up resistors for the comparison output signals. (Refer to Figure B-5 and Table B-3 for the location of the pull-up resistors for the comparison output signals.
Procedure 4339B EXT DCV2 EXT TRIG KEY LOCK HV OFF SW1-1 SW1-2 SW1-3 SW1-4 5 to 6 V Close Close Close Close 6 to 9 V Close Open Close Close 9 to 15 V Open Close Open Open 9. Reinstall the A1 main board, the A2 board, and the cover.
C Overload/Over-Current/No-Contact Operations Table C-1 shows the summary of operations when the 4339B detects OVLD (Overload), Over-Current (exceeding current limit), or N.C. (No-Contact). (Measurement parameter: R) Table C-1. OVLD/Over-Current/N.C. Operations (Meas. parameter: R) Display Data Comprtr Handler Output GPIB Output1 Mode Mode OVLD (Overload) OVLD LOW Over-Current OV CURR LOW N.C. N.C. (No-Contact) OVLD & OVLD N.C. OVLD & OV CURR Over-Current N.C. LOW LOW 2 : 1 /LO : 9.
Overload/Over-Current/No-Contact Operations 4339B When the measurement parameter is I, High or Low status in the operation of handler/GPIB output and display is reversed as shown in Table C-2. Table C-2. OVLD/Over-Current/N.C. Operations(Meas. parameter: I) Display Data Comprtr Handler Output GPIB Output1 Mode Mode OVLD (Overload) OVLD HIGH Over-Current OV CURR HIGH N.C. N.C. (No-Contact) OVLD & OVLD N.C. OVLD & OV CURR Over-Current N.C. HIGH HIGH 2 : 1 /HI : 9.
Error Messages This section lists the messages that are displayed on the 4339B's LCD display or transmitted by the instrument over GPIB, in numerical order.
Instrument Errors 4339B Out of Range OVER CURRENT The current limit is working and the source output is reduced. Lower the test voltage value, or raise the current limit value. OVLD The measured value is out of the measurable range. Change the measurement range appropriately, or use the Auto range mode. Warning WARNING: FIXTURE LEAKAGE When an OPEN correction is performed, the leakage current is high but it is less than the correction limit.
4339B 16 Instrument Errors PREV. SETTING LOST Instrument settings in the backup memory have been lost. The instrument keeps instrument settings in backup memory for 72 hours after being turned OFF. When this error occurs during power-on test, the 4339B continues operation and does not assert the /ALARM signal on the handler interface. When this error occurs during self-test, the 4339B stops operation and asserts the /ALARM signal. 17 SAVE FAILED The instrument setting was not be saved.
GPIB Errors 34 4339B HIGH OFFSET LEAKAGE When the calibration or the o set-error canceling is performed, the o set-error is too high. Contact your nearest Agilent Technologies o ce. 35 FIXTURE UNKNOWN An unknown test xture or test leads is connected to the Interlock connector; for example, when you connect the 16117C Test Leads to the 4339B, the wiring of the interlock cable is mistaken. 36 INTERLOCK OPEN Nothing is connected to the Interlock connector or the test xture's cover is opened.
4339B -109 GPIB Errors Missing parameter Fewer parameters were received than required for the header; for example, the [:SENSe]:AVERag:COUNt command requires one parameter, so receiving only :AVER:COUN is not allowed. -112 Program mnemonic too long The header contains more than twelve characters (see IEEE 488.2, 7.6.1.4.1). -113 Unde ned header The header is syntactically correct, but it is unde ned for the 4339B; for example, *XYZ is not de ned for the 4339B.
GPIB Errors -148 4339B Character data not allowed A legal character data element was encountered that's prohibited by the 4339B. -150 String data error This error, as well as errors 0151 and 0158, are generated when analyzing the syntax of a string data element. This particular error message is used if the 4339B cannot detect a more speci c error. -151 Invalid string data A string data element was expected, but was invalid for some reason (see IEEE 488.2, 7.7.5.
4339B -211 GPIB Errors Trigger ignored A GET, *TRG, or triggering signal was received and recognized by the 4339B but was ignored because of 4339B timing considerations; for example, the 4339B was not ready to respond. -213 Init ignored A request for a measurement initiation was ignored as another measurement was already in progress. -221 Settings con ict A legal program data element was analyzed syntax but could not be executed due to the current device state (see IEEE 488.2, 6.4.5.3 and 11.5.1.1.
GPIB Errors -400 4339B Query errors This is the generic query error that the 4339B cannot detect more speci c errors. This code indicates only that a query error as de ned in IEEE 488.2, 11.5.1.1.7 and 6.3 has occurred. -410 Query INTERRUPTED A condition causing an interrupted query error occurred (see IEEE 488.2, 6.3.2.3); for example, a query followed by DAB or GET before a response was completely sent. -420 Query UNTERMINATED A condition causing an unterminated query error occurred (see IEEE 488.
Index A :ABORt, 5-7 ABORt, 4-11 ABORt command, 5-7 Adrs key, 3-11 [:ALL?] CALibration , 5-16 ammeter o set voltage and input resistance test , 9-6 annunciator, 3-2 ARM, 5-46 DELay, 5-46 SOURce, 5-46 ARM event detection state, 5-46 :ARM[:SEQuence1][:LAYer]:DELay, 4-25 :ARM[:SEQuence1][:LAYer]:SOURce, 4-25 ARM subsystem , 5-8 Arrow key, 3-13 ASCII, 5-22, 5-48 :AUTO CALibration , 5-16 SENSe:CURRent:RANGe , 5-27 Auto/Hold key , 3-10 Auto range, 3-10 :AVERage:COUNt SENSe, 5-25 Average key , 3-5 :AVERage[:STATe]
how to use, 2-5, 4-9 limit, 3-12 on/o , 3-14 Con g key , 3-18 contact check, 3-15 how to perform, 4-8 how to set, 2-12 contact check , 5-26 Contact Check Fail, C-1 contact check functional test , 9-24 :CONTact:DATA? SENSe, 5-26 :CONTact:LIMit? SENSe, 5-26 :CONTact:OFFSet SENSe, 5-26 :CONTact:VERify SENSe, 5-26 Cont Chk key , 3-15 :CONTrol DATA:FEED, 5-18 :CORRection:COLLect[:ACQuire] SENSe, 5-26 :CORRection:DATA? SENSe, 5-27 :CORRection[:STATe] SENSe, 5-27 :COUNt SENSe:AVERage, 5-25 TRIGger:[:SEQuence1], 5-
:FORMat CALCulate1, 5-11 CALCulate3, 5-14 :FORMat[:DATA], 5-22 FORMat subsystem, 5-22 front panel, 3-2 :FUNCtion SENSe, 5-28 G GET, 4-11 :GLENgth CALCulate1:RESistivity, 5-14 GPIB address, 3-11 how to set, 2-19 group execution trigger, 4-11 H handler interface functional test , 9-22 High Voltage indicator , 3-4 Hold range, 3-10 I Idle state, 5-46 *IDN?, 5-37 IEEE 754 oating point format, 5-22, 5-49 IMON, 5-18 :INITiate:CONTinuous, 4-11, 5-23 :INITiate[:IMMediate], 4-11, 5-23 Initiate state, 5-46 INITi
measurement time mode, 3-5 GPIB command, 5-27 how to select, 2-2, 4-6 Min key , 3-13 Q N R N.C.
SENSe subsystem , 5-25 , 5-28 , 5-4 Seq Abort key , 3-11 Seq Mode key, 3-6 :SEQuence1 ARM, 5-9 [:SEQuence1]:COUNt TRIGger, 5-34 [:SEQuence1]:DELay TRIGger, 5-34 [:SEQuence1][:IMMediate] TRIGger, 5-34 [:SEQuence1][:LAYer]:DELay ARM, 5-9 [:SEQuence1][:LAYer]:SOURce ARM, 5-9 [:SEQuence1]:SOURce TRIGger, 5-35 [:SEQuence1]:TIMer TRIGger, 5-36 Sequence operation state, 5-47 serial number, A-2 Service Request Enable Register, 5-40 Service Request (SRQ), 5-40 Show Setting Key, 3-10
*TST?, 4-26, 5-39 U UNKNOWN terminals , 3-4 :UPPer CALCulate1:LIMit, 5-12 V :VERSion? SYSTem, 5-33 :VOLTage SOURce, 5-29 volume resistivity, 7-3 V Output indicator , 3-5 V Output key , 3-4 W *WAI, 5-39 Index-6
REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative. 11/29/99 United States: Agilent Technologies Test and Measurement Call Center P.O.