OVLenable <mask> OVLenable? <mask> ::= 8-bit integer in NR1 format as shown, 54640-series
only:
Bit Weight Input
4 16 External Trigger
3 8 Channel 4
2 4 Channel 3
1 2 Channel 2
0 1 Channel 1
n/a OVLRegister? <value> ::= integer in NR1 format, 54640-series only. See
OVLenable for <value>
*PMC n/a n/a
:POD<n>:DISPlay {{0 | OFF} | {1 | ON}} :POD<n>:DISPlay? {0 | 1}
<n> ::= 1-2 in NR1 format
:POD<n>:THReshold <type>[suffix] :POD<n>:THReshold? <n> ::= 1-2 in NR1 format
<type> ::= {CMOS | ECL | TTL | <user defined value>}
<user defined value> ::= value in NR3 format
[suffix] ::= {V | mV | uV }
:PRINt [parameter][,parameter]] :PRIN? [parameter][,parameter] x
4]
<parameter> ::= {HIRes | LORes | TIFF | PCL | BMP | PARallel |
DISK | FACTors | NOFactors}
<parameter> can be repeated up to 5 times.
*RCL <value> n/a <value> ::= {0 | 1 | 2}
*RST n/a See reset values in the online Programmer’s Reference.
:RUN n/a n/a
*SAV <value> n/a <value> ::= {0 | 1 | 2}
:SERial? <return value> ::= unquoted string containing serial number
:SINGle n/a n/a
*SRE <mask> *SRE? <mask> ::= sum of all bits that are set, 0 to 255; an integer in NR1
format. <mask> ::= following values:
Bit Weight Enables
7 128 OPER - Operation Status Register
664 Not Used
5 32 ESB - Event Status Bit
4 16 MAV - Message Available
3 8 Not used
2 4 MSG - Message
1 2 USR - User
0 1 TRG - Trigger
n/a :STATus? <display> {0 | 1}
<display> ::= {CHANnel<n> | DIGital0,...,DIGital15 | FUNCtion
|MATH};
<n> ::= 1-2 or 1-4 in NR1 format
Command Query Options and Query Returns