User`s guide
8-10
Programmer’s Quick Reference
Commands and Queries
:MEASure:VBASe [<source>] :MEASure:VBASe? [<source>] <source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<base_voltage> ::= voltage at the base of the selected waveform
in NR3 format
:MEASure:VMAX [<source>] :MEASure:VMAX? [<source>] <source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= maximum voltage of the selected waveform in
NR3 format
:MEASure:VMIN [<source>] :MEASure:VMIN? [<source>] <source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= minimum voltage of the selected waveform in
NR3 format
:MEASure:VPP [<source>] :MEASure:VPP? [<source>] <source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= voltage peak-to-peak of the selected
waveform in NR3 format
:MEASure:VRMS [<source>] :MEASure:VRMS? [<source>] <source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= calculated dc RMS voltage in NR3 format
n/a :MEASure:VTIMe?
<vtime>[,<source>]
<vtime> ::= displayed time from trigger in seconds in NR3 format
<return_value> ::= voltage at the specified time in NR3 format
<source> ::= {CHANnel<n> | FUNCtion | MATH} for 546xxA
<source> ::= {CHANnel<n> | DIGital0,...,DIGital15 | FUNCtion |
MATH} for 546xxD
<n> ::= 1-2 or 1-4 in NR1 format
:MEASure:VTOP [<source>] :MEASure:VTOP? [<source>] <source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= voltage at the top of the waveform in NR3
format
:MEASure:XMAX [<source>] :MEASure:XMAX? [<source>] <source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= horizontal value of the maximum in NR3 format
:MEASure:XMIN [<source>] :MEASure:XMIN? [<source>] <source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= horizontal value of the maximum in NR3 format
:MERGe <pixel memory> n/a <pixel memory> ::= {PMEMory{0 | 1 | 2}}
*OPC *OPC? ASCII ""1"" is placed in the output queue when all pending device
operations have completed.
:OPEE <n> :OPEE? <n> ::= 16-bit integer in NR1 format
n/a :OPER? <n> ::= 16-bit integer in NR1 format
n/a *OPT? <return_value> ::= n, A.XX.XX
n identifies the module.
XX.XX identifies the module software revision.
N2757A, A.XX.XX
Command Query Options and Query Returns