Specifications
3-17
Testing Performance
To verify trigger sensitivity
To verify trigger sensitivity
This test verifies the trigger sensitivity. In this test, you will apply 25 MHz to the
oscilloscope. You will then decrease the amplitude of the signal to the specified
levels, and check to see if the oscilloscope is still triggered. You will then repeat
the process at the upper bandwidth limit.
Test limits for the Internal trigger:
<10 mV/div (dc to max bandwidth): greater of 1 div or 5 mV
p-p
≥10 mV/div (dc to max bandwidth): 0.6 div
Test limits for the External trigger:
dc to 100 MHz: <100 mV
p-p
100 MHz to max bandwidth: <200 mV
p-p
Table 3-9 Equipment Required to Verify Trigger Sensitivity
Equipment Critical Specifications Recommended Model/Part
Oscilloscope Calibrator 25-MHz, 100-MHz, 350-MHz and
500-MHz sine waves
Fluke 5820A
Power splitter Outputs differ < 0.15 dB Agilent 11667B
Cable * BNC, Qty 3 Agilent 10503A
Adapter N (m) to BNC (f), Qty 3 Agilent 1250-0780
* The oscilloscope calibrator is supplied with 2 or more coaxial cables N (m), BNC (m),
1 meter long, Fluke P/N 686318.