Specifications

97
Chapter 3: Testing Performance
To Test the Single-clock, Multiple-edge, State Acquisition
If the Label selection field reads Label1_TZ, you must select Label1 for the search
term. Follow the same procedure as in b above.
d In the Marker Setup window, select the ‘occurs’ value field that
corresponds to marker G1. Enter 4095.
e In the Marker Setup window, select the ‘occurs’ value field that
corresponds to marker G2. Enter 4096.
f Select Close to apply the marker values to the data. If the “Pattern NOT
found for marker...” error message does not appear, then the test
passes. Record the Pass or Fail in the performance test record.
5 Repeat steps 3 and 4 for the next clock edge listed in the table in step 3,
until all listed clock edges have been tested.
6 If the setup/hold used for the previous steps was 5.0/-2.0 ns, repeat steps 1
through 5 using setup/hold -1.5/4.5 ns. If the setup/hold used for the
previous steps was -1.5/4.5 ns, continue on with the next section.