Specifications

84
Chapter 3: Testing Performance
To Test the Single-clock, Multiple-edge, State Acquisition
Set up the logic analyzer
Perform the following steps if you have not done so for the previous tests.
1 Set up the Sampling tab.
a In the Analyzer window, select the Sampling tab.
b Select State Mode.
2 Assign all pods to Analyzer 1.
a In the Analyzer setup window, select the Format tab.
b Under the Format tab, select Pod Assignment.
c In the Pod Assignment window, use the mouse to drag the pods to the
Analyzer 1 column.
d Select Close to close the Pod Assignment window.