Specifications
80
Chapter 3: Testing Performance
To Test the Multiple-clock, Multiple-edge, State Acquisition
If the Label selection field reads Label1_TZ, you must select Label1 for the search
term. Follow the same procedure as in b above.
d In the Marker Setup window, select the ‘occurs’ value field that
corresponds to marker G1. Enter 4095.
e In the Marker Setup window, select the ‘occurs’ value field that
corresponds to marker G2. Enter 4096.
f Select Close to apply the marker values to the data. If the “Pattern NOT
found for marker...” error message does not appear, then the test
passes. Record the Pass or Fail in the performance test record.
6 Repeat steps 4 and 5 for the next clock edge combination listed in the
table in step 4, until both clock combinations have been tested.
7 Enable the pulse generator channel 1 COMP (LED on).