Specifications

79
Chapter 3: Testing Performance
To Test the Multiple-clock, Multiple-edge, State Acquisition
5 Verify the test data.
a In the Listing window, select Run. The display should show an
alternating pattern of “AA” and “55”.
b In the Marker Setup window, select the Define... field associated with
G1, and the G1 Marker Pattern window appears. In the pattern field,
enter “AA”. Select Apply, then select Close.
If the label selection field reads Label1_TZ, you must select Label1 for the search
term. To do this, select Label1_TZ; then, in the popup menu, select Replace label. In
the Replace popup menu, select Label1, then Apply, then Close.
c In the Marker Setup window, select the Define... field associated with
G2, and the G2 Marker Pattern window appears. In the pattern field,
enter “55”. Select Apply, then select Close.