To Test the Single-clock, Single-edge, State Acquisition
9 Select the clock to be tested.
a In the Analyzer setup window, select the Sampling tab.
b Under the Sampling tab, select the clock edge field under the clock to
be tested. Then select Falling Edge. The first time through this test,
select the first clock and edge. Ensure all other clocks are turned off.
Clocks
c Connect the clock input channel to be tested to the pulse generator
channel 1 OUTPUT. Disconnect all other clock input channels.
10 Verify the test data.
a In the Listing window, select Run. The display should show an
alternating pattern of “AA” and “55”.
b If the “Pattern NOT found for marker...” error message does not appear,
then the test passes. Record the Pass or Fail in the performance test
record.
11 Repeat steps 9 and 10 for the next clock edge listed in the table in step 9,
until all listed clock edges have been tested.