Specifications

60
Chapter 3: Testing Performance
To Test the Single-clock, Single-edge, State Acquisition
4 Select the clock to be tested.
a In the Analyzer setup window, select the Sampling tab.
b Under the Sampling tab, select the clock edge field under the clock to
be tested. Then select Rising Edge. Turn off all other clocks. The first
time through this test, select the first clock and edge.
Clocks
c Connect the clock input channel to be tested to the pulse generator
channel 1 OUTPUT. Disconnect all other clock input channels.
- . / 0