Specifications

43
Chapter 3: Testing Performance
To Test the Threshold Accuracy
Test the ECL threshold
1 In the Pod Threshold window, select the Standard threshold voltage field.
At the pop-up menu select ECL (–1.30 V).
2 On the function generator front panel, enter -1.214 V ±1 mV DC offset.
Use the multimeter to verify the voltage.
The activity indicators for Pod 1 should show all data channels and the J-clock
channel at a logic high.
Threshold Field