Specifications
32
Chapter 3: Testing Performance
To ensure the logic analyzer is operating as specified, software tests (self-tests)
and manual performance tests are done. The logic analyzer is considered
performance-verified if all of the software tests and manual performance tests
have passed. The procedures in this chapter indicate what constitutes a “Pass”
status for each of the tests.
Test Strategy
This chapter shows the module being tested in an Agilent Technologies 16700B-
series mainframe. For a complete test, start at the beginning with the software
tests and continue through to the end of the chapter. For an individual test,
follow the procedure in the test.
One-card Module. To perform a complete test on a one-card module, start at
the beginning of the chapter and follow each procedure.
Multi-card Module. To perform a complete test on a multi-card module,
perform the self-tests with the cards connected. Then, remove the multi-card
module from the mainframe and configure each card as a one-card module.
Install the one-card modules into the mainframe and perform the one-card
manual performance verification tests on each card. When the tests are complete,
remove the one-card modules, reconfigure them into a multi-card module,
reinstall it into the mainframe and perform the final multi-card test. For removal
instructions, see Chapter 6, “Replacing Assemblies.” For installation and
configuration instructions, see Chapter 2, “Preparing for Use.”
Test Interval
Test the performance of the module against specifications at two-year intervals.
Test Record Description
A performance test record for recording the results of each procedure is located
at the end of this chapter. Use the performance test record to gauge the
performance of the module over time.
Test Equipment
Each procedure lists the recommended test equipment. You can use equipment
that satisfies the specifications given. However, the procedures are based on
using the recommended model or part number.
Instrument Warm-Up
Before testing the performance of the module, warm-up the instrument and the
test equipment for 30 minutes.