Chapter 6: Replacing Assemblies
5 Position all cards and filler panels so that the endplates overlap.
6 Seat the cards and tighten the thumbscrews.
Starting with the bottom card, firmly seat the cards into the backplane connector
of the mainframe. Keep applying pressure to the center of the card endplate
while tightening the thumbscrews finger-tight. Repeat this for all cards and filler
panels starting at the bottom and moving to the top.
CAUTION: Correct air circulation keeps the instrument from overheating. For correct air circulation,
filler panels must be installed in all unused card slots. Keep any extra filler panels for
future use.
To replace the probe cable
1 Remove power from the instrument.
a Exit all logic analysis sessions. In the session manager, select
Shutdown.
b At the query, select Power Down.
c When the “OK to power down” message appears, turn the instrument
off.
d Disconnect the power cord.
2 Remove the screws that hold the probe cable to the rear panel of the
module.
3 Remove the faulty probe cable from the connector and install the
replacement cable.
4 Install the label on the new probe.
If you order a new probe cable, you will need to order new labels. Probe cables
shipped with the module are labeled. Probe cables shipped separately are not
labeled. Refer to chapter 7, "Replaceable Parts," for the part numbers and