Chapter 6: Replacing Assemblies
CAUTION: Turn off the instrument before installing, removing, or replacing a module in the
instrument.
Tools Required
A T10 TORX screwdriver, to remove screws connecting the probe cables and screws
connecting the back panel.
To remove the module
CAUTION: Electrostatic discharge can damage electronic components. Use grounded wriststraps
and mats when performing any service to this module.
1 Remove power from the instrument.
a Exit all logic analysis sessions. In the session manager, select
Shutdown.
b At the query, select Power Down.
c When the “OK to power down” message appears, turn the instrument
off.
d Disconnect the power cord.
2 Loosen the thumb screws.
Starting from the top, loosen the thumb screws on the filler panels and cards
located above the module and the thumb screws of the module.
3 Starting from the top, pull the cards and filler panels located above the
module half-way out.
4 If the module consists of a single card, pull the card completely out.
If the module consists of multiple cards, pull all cards completely out.