Specifications
104
Chapter 3: Testing Performance
To Test the Time Interval Accuracy
e In the marker Setup window, select the Define... field associated with
G2, and the G2 Marker Pattern window appears. In the Pattern field,
enter “1”.
Select the Pattern Qualify field and select When Entering.
If the Label selection field reads Label1_TZ, you must select Label1 for the search
term. Follow the same procedure as in b above.
In the Marker Pattern window, select Apply, then select Close.
4 Acquire the data.
a In the Waveform window, move the mouse cursor over Run and click
and hold the right mouse button. At the pop-up menu, select
Repetitive.
b Select Run. The logic analyzer repetitively acquires data.
c Continuously observe the Interval Time from G1 to G2=value in the
Marker Setup window.
Allow the logic analyzer to run repetitively for approximately one minute. If the
Interval Time value remains inside the range 749.921
µs to 750.079 µs, the test passes.
Record a Pass or Fail in the performance test record.
d Select Stop to end the acquisition.