Specifications

103
Chapter 3: Testing Performance
To Test the Time Interval Accuracy
3 Configure the Markers to measure the time interval.
a In the Marker Setup window select the Time field associated with G1,
and select Pattern. Select the Time field associated with G2, and select
Pattern.
b Select the Occurs field associated with G1 and enter “1”. Select the
Occurs field associated with G2 and enter “30000”.
c Select the From field associated with G2 and select G1.
In the Marker Setup Window, you will observe the Interval Time from G1 to G2=value
to determine the pass or fail status of this test.
d In the marker Setup window, select the Define... field associated with
G1, and the G1 Marker Pattern window appears. In the Pattern field,
enter “1”.
Select the Pattern Qualify field and select When Entering.
If the label selection field reads Label1_TZ, you must select Label1 for the search
term. To do this, select Label1_TZ; then, in the popup menu, select Replace label. In
the Replace popup menu, select Label1, then Apply, then Close.
In the Marker Pattern window, select Apply, then select Close.