Adjustment
Null Loop Phase Correction Constants
1. Chapter Title
2. Chapter Title 3. Adjustment
4. Chapter Title
5. Chapter Title
Null Loop Phase Correction Constants
The purpose of this procedure is to adjust the phase of the null loop.
Required Equipment
Description Recommended Model
4TP OPEN termination 42090A
Procedure
1. Run the adjustment program.
2. Choose the Adj_NullPhase.
3. Connect the OPEN termination as shown in Figure 3-4
Figure 3-4 Null Loop Correction Constants Setup
4. Follow the adjustment program instruction to update the correction constants.