Service manual
3 89
Adjustment
LPOT-Mixer Offset Correction Constants
1. Chapter Title
2. Chapter Title 3. Adjustment
4. Chapter Title
5. Chapter Title
L
POT
-Mixer Offset Correction Constants
The purpose of this procedure is to adjust the L
POT
-Mixer Offset.
Required Equipment
None
Procedure
1. Run the adjustment program.
2. Choose the Adj_LpMix_Leak.
3. Confirm nothing is connected to the 4294A UNKOWN terminal.
4. Follow the adjustment program instruction to update the correction constants.