Service manual
88 3
Adjustment
Image Rejection Correction Constants
Image Rejection Correction Constants
The purpose of this procedure is to generate the correction constants that are used to adjust
the Vector Ratio Detector.
Required Equipment
None
Procedure
1. Run the adjustment program.
2. Choose the Adj_ImageRej.
3. Confirm nothing is connected to the 4294A UNKOWN terminal.
4. Follow the adjustment program instruction to update the correction constants.