Service manual
82 3
Adjustment
Safety Considerations
Safety Considerations
This manual contains NOTEs, CAUTIONs, and WARNINGs that must be followed to
ensure the safety of the operator and to keep the instrument in a safe and serviceable
condition. The Adjustment must be performed by qualified service personnel.
WARNING Any interruption of the protective ground conductor (inside or outside the analyzer)
or disconnection of the protective ground terminal can make the can make the
instrument dangerous. Intentional interruption of the protective ground system for
any reason is prohibited.
Required Controller
The following controller system is required to run the adjustment program.
Windows PC PC-AT Compatible, RAM:≥64MBytes, CPU Pentium 200 MHz or
faster
OS Microsoft
®
Windows XP
®
, Windows 2000
®
Software HP VEE (6.1 or 6.2)
GPIB Card 82350, 82340, 82341
Required Equipment
Table 1-1 lists the equipment required to perform the Adjustment procedures described in
this chapter. Use only calibrated test equipment when adjusting the 4294A.
Warm-up for Adjustment
Warm-up the analyzer for at least 30 minute before performing any of the following
Adjustment procedures to ensure procedures to ensure proper results and correct
instrument operation.