Performance Test
Performance Test Record
Accessory: 2m Test Leads
Standard: 100 pF
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 1 kHz Cp ±0.49 pF pF ±0.010 pF
D ±0.0049 ±0.0001
100 kHz Cp ±0.104 pF pF ±0.025 pF
D ±0.0010 ±0.0002
1 MHz Cp ±0.128 pF pF ±0.025 pF
D ±0.0013 ±0.0002
3 MHz Cp ±0.42 pF pF ±0.030 pF
D ±0.0042 ±0.0005
10 MHz Cp ±1.33 pF pF ±0.15 pF
D ±0.0133 ±0.003
Accessory: 2m Test Leads
Standard: 1 Ω
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 1 kHz R ±0.0055 Ω Ω ±0.001 Ω
Accessory: 2m Test Leads
Standard: 100 mΩ
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 1 kHz R ±4.6 mΩ mΩ ±0.1 mΩ