Service manual
2 79
Performance Test
Performance Test Record
2. Performance Test
Accessory: 1m Test Leads
Standard: 1 Ω
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 1 kHz R ±0.0055 Ω Ω ±0.001 Ω
Accessory: 1m Test Leads
Standard: 100 mΩ
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 1 kHz R ±4.6 mΩ mΩ ±0.1 mΩ
Accessory: 2m Test Leads
Standard: 10 pF
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 1 kHz Cp ±0.41 pF pF ±0.0010 pF
D ±0.041 ±0.0001
1 MHz Cp ±0.015 pF pF ±0.0025 pF
D ±0.0015 ±0.0002
3 MHz Cp ±0.045 pF pF ±0.0030 pF
D ±0.0045 ±0.0005
10 MHz Cp ±0.134 pF pF ±0.015 pF
D ±0.0134 ±0.003