Service manual
78 2
Performance Test
Performance Test Record
Accessory: 1m Test Leads
Standard: 10 pF
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 1 kHz Cp ±0.41 pF pF ±0.0010 pF
D ±0.041 ±0.0001
1 MHz Cp ±0.013 pF pF ±0.0025 pF
D ±0.0013 ±0.0002
3 MHz Cp ±0.041 pF pF ±0.0030 pF
D ±0.0041 ±0.0005
10 MHz Cp ±0.124 pF pF ±0.015 pF
D ±0.0124 ±0.003
Accessory: 1m Test Leads
Standard: 100 pF
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 1 kHz Cp ±0.49 pF pF ±0.010 pF
D ±0.0049 ±0.0001
100 kHz Cp ±0.102 pF pF ±0.025 pF
D ±0.0010 ±0.0002
1 MHz Cp ±0.128 pF pF ±0.025 pF
D ±0.0013 ±0.0002
3 MHz Cp ±0.38 pF pF ±0.030 pF
D ±0.0038 ±0.0005
10 MHz Cp ±1.22 pF pF ±0.15 pF
D ±0.0122 ±0.003