Service manual

2 75
Performance Test
Performance Test Record
2. Performance Test
Accessory: Nothing
Standard: 0.1 μF
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 120 Hz Cp ±0.30 nF nF ±0.025 nF
D ±0.0030 ±0.0001
1 kHz Cp ±0.076 nF nF ±0.01 nF
D ±0.0008 ±0.0001
10 kHz Cp ±0.077 nF nF ±0.025 nF
D ±0.0008 ±0.0002
0.1 V 1 kHz Cp ±0.085 nF nF ±0.01 nF
D ±0.0008 ±0.0001
Accessory: Nothing
Standard: 1 μF
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 120 Hz Cp ±0.0030 μF μF ±0.0004 μF
D ±0.0030 ±0.00015
1 kHz Cp ±0.00077 μF μF ±0.0001 μF
D ±0.0008 ±0.0001
0.1 V 1 kHz Cp ±0.00085 μF μF ±0.0001 μF
D ±0.0008 ±0.0001