Service manual
74 2
Performance Test
Performance Test Record
Accessory: Nothing
Standard: 1000 pF
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 1 kHz Cp ±1.15 pF pF ±0.10 pF
D ±0.0012 ±0.0001
10 kHz Cp ±0.79 pF pF ±0.10 pF
D ±0.0008 ±0.0001
100 kHz Cp ±0.76 pF pF ±0.10 pF
D ±0.0008 ±0.0001
0.1 V 1 kHz Cp ±2.8 pF pF ±0.10 pF
D ±0.0028 ±0.0001
Accessory: Nothing
Standard: 0.01 μF
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 120 Hz Cp ±0.033 nF nF ±0.0025 nF
D ±0.0033 ±0.0001
1 kHz Cp ±0.0079 nF nF ±0.001 nF
D ±0.0008 ±0.0001
10 kHz Cp ±0.0076 nF nF ±0.0025 nF
D ±0.0008 ±0.0002
0.1 V 1 kHz Cp ±0.0102 nF nF ±0.001 nF
D ±0.0010 ±0.0001