Service manual
○
C
Copyright 2009 Agilent Technologies
Accessory: 2m Test Leads
Standard: 10 pF
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V
1 kHz
Cp ± 0.41 pF pF ± 0.0010 pF 46.6
D ± 0.041 ± 0.0001 81.6
1 MHz
Cp ± 0.015 pF pF ± 0.0025 pF 5.0
D ± 0.0015 ± 0.0002 5.1
3 MHz
Cp ± 0.045 pF pF ± 0.0030 pF 13.0
D ± 0.0045 ± 0.0005 15.5
10 MHz
Cp ± 0.134 pF pF ± 0.015 pF 29.5
D ± 0.0134 ± 0.003 10.6
Accessory: 2m Test Leads
Standard: 100 pF
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V
1 kHz
Cp ± 0.49 pF pF ± 0.010 pF 14.1
D ± 0.0049 ± 0.0001 12.2
100 kHz
Cp ± 0.104 pF pF ± 0.025 pF 3.4
D ± 0.0010 ± 0.0002 3.6
1 MHz
Cp ± 0.128 pF pF ± 0.025 pF 4.3
D ± 0.0013 ± 0.0002 4.4
3 MHz
Cp ± 0.42 pF pF ± 0.030 pF 10.0
D ± 0.0042 ± 0.0005 13.2
10 MHz
Cp ± 1.33 pF pF ± 0.15 pF 6.0
D ± 0.0133 ± 0.003 31.5
Accessory: 2m Test Leads
Standard: 1 Ω
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V 1 kHz R
± 0.0055 Ω Ω ± 0.001 Ω
16.4
Accessory: 2m Test Leads
Standard: 100 mΩ
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V 1 kHz R
± 4.6 mΩ mΩ ± 0.1 mΩ
22.4