Service manual
2 71
Performance Test
Performance Test Record
2. Performance Test
DC Bias Level Accuracy Test
Bias Level Test Limit Test R e sul t Measurement
Uncertainty
0 V ±5.0 mV mV ±0.001 mV
25 V ±30.3 mV mV ±0.61 mV
40 V ±45.6 mV mV ±0.86 mV
–25 V ±30.3 mV mV ±0.61 mV
–40 V ±45.6 mV mV ±0.86 mV
0 mA ±0.20 mA mA ±0.001 mA
20 mA ±0.60 mA mA ±0.004 mA
100 mA ±2.20 mA mA ±0.016 mA
–20 mA ±0.60 mA mA ±0.004 mA
–100 mA ±2.20 mA mA ±0.016 mA