Service manual

70 2
Performance Test
Performance Test Record
OSC Level Accuracy (with 42942A Terminal Adapter)
OSC Level Frequency Test Li mi t Test Result Measurement
Uncertainty
0.5 V 1 MHz ±51.3 mV mV ±5.2 mV
10 MHz ±53.5 mV mV ±5.3 mV
15 MHz ±54.8 mV mV ±5.3 mV
16 MHz ±55.0 mV mV ±5.3 mV
50 MHz ±63.5 mV mV ±5.4 mV
110 MHz ±78.5 mV mV ±5.7 mV
DC Bias Monitor Accuracy Test
Bias Level Test Limit Test R e sul t Measurement
Uncertainty
0 V ±5.0 mV mV ±0.001 mV
25 V ±55.0 mV mV ±0.61 mV
40 V ±85.0 mV mV ±0.86 mV
–25 V ±55.0 mV mV ±0.61 mV
–40 V ±85.0 mV mV ±0.86 mV
0 mA ±0.50 mA mA ±0.001 mA
20 mA ±0.75 mA mA ±0.004 mA
100 mA ±1.75 mA mA ±0.016 mA
–20 mA ±0.75 mA mA ±0.004 mA
–100 mA ±1.75 mA mA ±0.016 mA