Service manual

C
Copyright 2009 Agilent Technologies
Accessory: 42942A Terminal Adapter
Standard: 10 cm Airline with SHORT
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V
1 MHz
Z ± 0.0062 Ω Ω ± 0.0017 Ω
3.5
θ
± 54.4 mrad mrad ± 2.0 mrad 27.2
10 MHz
Z ± 0.0165 Ω Ω ± 0.0043 Ω
2.0
θ
± 15.3 mrad mrad ± 4.0 mrad 3.8
100 MHz
Z ± 0.157 Ω Ω ± 0.027 Ω
1.5
θ
± 14.7 mrad mrad ± 2.5 mrad 5.9
Accessory: 1m Test Leads
Standard: 10 pF
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V
1 kHz
Cp ± 0.41 pF pF ± 0.0010 pF 61.9
D ± 0.041 ± 0.0001 53.1
1 MHz
Cp ± 0.013 pF pF ± 0.0025 pF 4.5
D ± 0.0013 ± 0.0002 4.5
3 MHz
Cp ± 0.041 pF pF ± 0.0030 pF 14.2
D ± 0.0041 ± 0.0005 14.3
10 MHz
Cp ± 0.124 pF pF ± 0.0010 pF 39.3
D ± 0.0124 ± 0.003 32.7
Accessory: 1m Test Leads
Standard: 100 pF
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V
1 kHz
Cp ± 0.49 pF pF ± 0.010 pF 16.5
D ± 0.0049 ± 0.0001 15.5
100 kHz
Cp ± 0.102 pF pF ± 0.025 pF 3.5
D ± 0.0010 ± 0.0002 3.5
1 MHz
Cp ± 0.128 pF pF ± 0.025 pF 4.0
D ± 0.0013 ± 0.0002 4.0
3 MHz
Cp ± 0.38 pF pF ± 0.0030 pF 12.9
D ± 0.0038 ± 0.0005 13.2
10 MHz
Cp ± 1.22 pF pF ± 0.15 pF 27.2
D ± 0.0122 ± 0.003 31.4
Accessory: 1m Test Leads
Standard: 1 Ω
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V 1 kHz R
± 0.0055 Ω Ω ± 0.001Ω
21.7
Accessory: 1m Test Leads
Standard: 100 mΩ
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V 1 kHz R
± 4.6 mΩ mΩ ± 0.1 mΩ
19.2