Service manual

2 53
Performance Test
Calculation Sheet
2. Performance Test
OSC Level Accuracy (Accessory: Nothing, 1 MHz to 110 MHz)
OSC
Level
[a]
Freq. Monitor
Reading
[b]
Power
Meter
Reading
[c]
Calculated
OSC level
[d]
250 c×=
Monitor
Accuracy
[b–d]
Monitor
Accuracy
Limit
OPEN
Reading
[e]
Test Result
[e×d/b–a]
0.5 V 1 MHz V W V V ± 0.0904 V V mV
10 MHz V W V V ± 0.0938 V V mV
15 MHz V W V V ± 0.0956 V V mV
16 MHz V W V V ± 0.0960 V V mV
50 MHz V W V V ± 0.1088 V V mV
110 MHz V W V V ± 0.1313 V V mV
OSC Level Accuracy (with 42942A Terminal Adapter)
OSC
Level
[a]
Freq. Monitor
Reading
[b]
Power
Meter
Reading
[c]
Calculated
OSC level
[d]
250 c×=
Monitor
Accuracy
[b–d]
Monitor
Accuracy
Limit
OPEN
Reading
[e]
Test Result
[e×d/b–a]
0.5 V 1 MHz V W V V ± 0.0904 V V mV
10 MHz V W V V ± 0.0938 V V mV
15 MHz V W V V ± 0.0956 V V mV
16 MHz V W V V ± 0.0960 V V mV
50 MHz V W V V ± 0.1088 V V mV
110 MHz V W V V ± 0.1313 V V mV