Service manual
○
C
Copyright 2009 Agilent Technologies
Accessory: Nothing
Standard: 10 Ω
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V 1 kHz R
± 0.0100 Ω Ω ± 0.003 Ω
4.9
0.1 V 1 kHz R
± 0.0114 Ω Ω ± 0.003 Ω
6.3
Accessory: Nothing
Standard: 1 Ω
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V 1 kHz R
± 0.0033 Ω Ω ± 0.001 Ω
13.6
0.1 V 1 kHz R
± 0.0040 Ω Ω ± 0.001 Ω
14.7
Accessory: Nothing
Standard: 100 mΩ
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V 1 kHz R
± 2.6 mΩ mΩ ± 0.1 mΩ
13.0
0.1 V 1 kHz R
± 3.2 mΩ mΩ ± 0.1 mΩ
31.0
Accessory: Nothing
Standard: 10 mΩ
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V 1 kHz R
± 2.5 mΩ mΩ ± 0.01 mΩ
10.2
0.1 V 1 kHz R
± 3.1 mΩ mΩ ± 0.01 mΩ
38.9
Accessory: 42942A Terminal Adapter
Standard: 50 Ω
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V
1 kHz
⏐Z⏐ ± 0.31 Ω Ω ± 0.090 Ω
2.8
θ
± 6.1 mrad mrad ± 1.8 mrad 3.4
1 MHz
⏐Z⏐ ± 0.31 Ω Ω ± 0.090 Ω
2.8
θ
± 6.1 mrad mrad ± 1.8 mrad 3.4
10 MHz
⏐Z⏐ ± 0.31 Ω Ω ± 0.090 Ω
2.8
θ
± 6.2 mrad mrad ± 1.8 mrad 3.4
100 MHz
⏐Z⏐ ± 0.54 Ω Ω ± 0.010 Ω
4.0
θ
± 10.9 mrad mrad ± 2.0 mrad 5.5
Accessory: 42942A Terminal Adapter
Standard: 10 cm Airline with OPEN
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V
1 MHz
⏐Z⏐ ± 155 Ω Ω ± 44 Ω
4.1
θ
± 7.1 mrad mrad ± 2.0 mrad 3.6
10 MHz
⏐Z⏐ ± 15.7 Ω Ω ± 4.4 Ω
3.7
θ
± 7.2 mrad mrad ± 2.0 mrad 3.6
100 MHz
⏐Z⏐ ± 2.32 Ω Ω ± 0.43 Ω
7.3
θ
± 10.8 mrad mrad ± 2.0 mrad 5.4