Service manual
Table 2-14 Measurement Accuracy Test(Accesory:2m Test Leads,
Standard Capacitor) Setting
Standard
Capacitor
OSC Level Frequency
10 pF 0.5 V 1 kHz
1 MHz
3 MHz
10 MHz
100 pF 0.5 V 1 kHz
100 kHz
1 MHz
3 MHz
10 MHz
2 51
Performance Test
Performance Test
2. Performance Test
56. Connect the Standard resistor as shown in Figure 2-18.
57. Press [Meas] - R-X to set the measurement mode to R-X.
58. Press [Trigger] - Single to make a measurement.
59. Record the 4294A reading on the calculation sheet for all setting in Table 2-15, then
Calculate the test result.
Table 2-15 Measurement Accuracy Test(Accesory:2m Test Leads,
Standard Resistor) Setting
Standard
Capacitor
OSC Level Frequency
1 Ω 0.5 V 1 kHz
100 mΩ 0.5 V 1 kHz