Service manual
2 47
Performance Test
Performance Test
2. Performance Test
Figure 2-15 Connection for phase compensation
d. Press PHASE COMP to measure the phase compensation data.
e. Connect the 100 Ω Resistor’s all terminals to the test leads as shown in Figure 2-16.
Figure 2-16 Connection for load compensation
f. Press LOAD to measure the load compensation data.
g. Press done to activate the compensation data.