Service manual

2 45
Performance Test
Performance Test
2. Performance Test
c. Gently inserts the airline center conductor into the open termination center
conductor. Mate the outer conductors. Then tie the connection by the 136 N·cm
torque.
Figure 2-13 10cm Airline with OPEN Measurement Test Setup
29. Press [Trigger] - Single to make a measurement.
30. Record the 4294A reading on the calculation sheet for all setting in Table 2-11, then
calculate the test result.
Table 2-11 Measurement Accuracy Test (Accessory: 42942A
Terminal Adapter, SHORT or OPEN Termination)
Setting
OSC Level Frequency
0.5 V 1 MHz
10 MHz
100 MHz
31. Connect the 10cm airline and SHORT termination to the 7mm connector as follows.
(refer to
Figure 2-14)
a. Remove the OPEN termination from the airline.
b. Gently inserts the airline center conductor into the short termination center
conductor. Mate the outer conductors. Then tie the connection by the 136 N·cm