Service manual

44 2
Performance Test
Performance Test
21. Press [Cal] - Adapter - 7mm 42942A to set the adapter to the 42942A.
22. Setup the 4294A as follows.
NOTE If Setup is completed in OSC level accuracy test, this operation is not required.
a. Press SETUP.
b. Connect the OPEN termination furnished with the 42942A to the 7mm connector.
c. Press PHASE COMP to measure the phase compensation data.
d. Press OPEN to measure the open compensation data.
e. Connect the SHORT termination furnished with the 42942A instead of the OPEN
termination.
f. Press SHORT to measure the short compensation data.
g. Connect the 50 Ω termination furnished with the 42942A instead of the SHORT
termination.
h. Press LOAD to measure the load compensation data.
i. Press done to activate the measured compensation data.
23. Press [Meas] - |Z|-θ to set the measurement mode.
24. Press [B] - [Format] - PHASE UNIT to set the phase display unit to radian.
25. Connect the 50 Ω Termination furnished with the 16190A to the 7mm connector.
26. Press [Trigger] - Single to make a measurement.
27. Record the 4294A reading on the calculation sheet for all setting in Table 2-10, then
calculate the test result.
Table 2-10 Measurement Accuracy Test (Accessory: 42942A
Terminal Adapter, 50 Ω) Setting
OSC Level Frequency
0.5 V 1 kHz
1 MHz
10 MHz
100 MHz
28. Connect the 10 cm airline and the OPEN termination to the 7mm connector as
follows.(refer to
Figure 2-13)
a. Fully retract the threads on the test head 7mm connector. Then insert the marked
side tip of the airline center conductor into the connector center conductor.
b. Gently cover the airline center conductor with the airline outer conductor, with the
Agilent logo side down. (To prevent damage, do not let the center conductor scrape
the edge of the outer conductor.) Mate the outer conductors. Then tie the connection
by the 136 N·cm. (A 1/2 inch open end wrench may be necessary to hold the airline
stationary)