Service manual

2 43
Performance Test
Performance Test
2. Performance Test
16. Connect the 10 Ω standard resistor to the 4294A UNKNOWN Terminal as shown in
Figure 2-9 on page 39.
17. Press [Meas] - R-X to set the measurement mode to R-X.
18. Press [Trigger] - SINGLE to make a measurement.
19. Record the 4294A reading on the calculation sheet for all setting in Table 2-9, then
calculate the test result.
Table 2-9 Measurement Accuracy Test(Accesory:None, Standard
Resister) Setting
Standard
Capacitor
OSC Level Frequency
10 Ω 0.5 V 1 kHz
0.1 V 1 kHz
1 Ω 0.5 V 1 kHz
0.1 V 1 kHz
100 mΩ 0.5 V 1 kHz
0.1 V 1 kHz
10 mΩ 0.5 V 1 kHz
0.1 V 1 kHz
20. Disconnect all the equipment from the 4294A, then connect the equipment as shown in
Figure 2-12.
Figure 2-12 Measurement Accuracy Test(Accessory: 42942A Terminal Adapter)