Service manual

C
Copyright 2009 Agilent Technologies
Accessory: Nothing
Standard: 1000 pF
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V
1 kHz
Cp ± 1.15 pF pF ± 0.10 pF 3.8
D ± 0.0012 ± 0.0001 3.9
10 kHz
Cp ± 0.79 pF pF ± 0.10 pF 2.7
D ± 0.0008 ± 0.0001 2.7
100 kHz
Cp ± 0.76 pF pF ± 0.10 pF 2.5
D ± 0.0008 ± 0.0001 2.5
0.1 V 1 kHz
Cp ± 2.8 pF pF ± 0.10 pF 9.2
D ± 0.0028 ± 0.0001 9.0
Accessory: Nothing
Standard: 0.01 μF
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V
120 Hz
Cp ± 0.033 nF nF ± 0.0025 nF 12.4
D ± 0.0033 ± 0.0001 12.5
1 kHz
Cp ± 0.0079 nF nF ± 0.001 nF 3.0
D ± 0.0008 ± 0.0001 3.0
10 kHz
Cp ± 0.0076 nF nF ± 0.0025 nF 2.9
D ± 0.0008 ± 0.0002 2.8
0.1 V 1 kHz
Cp ± 0.0102 nF nF ± 0.001 nF 3.8
D ± 0.0010 ± 0.0001 3.8
Accessory: Nothing
Standard: 0.1 μF
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V
120 Hz
Cp ± 0.30 nF nF ± 0.0025 nF 11.5
D ± 0.0030 ± 0.0001 11.3
1 kHz
Cp ± 0.076 nF nF ± 0.01 nF 2.8
D ± 0.0008 ± 0.0001 2.9
10 kHz
Cp ± 0.077 nF nF ± 0.025 nF 2.9
D ± 0.0008 ± 0.0002 2.8
0.1 V 1 kHz
Cp ± 0.085 nF nF ± 0.01 nF 3.1
D ± 0.0008 ± 0.0001 3.2
Accessory: Nothing
Standard: 1 μF
OSC
Level
Frequency Parameter Test Limits
Test
Result
Measurement
Uncertainty
TAR
0.5 V
120 Hz
Cp
± 0.0030 μF μF
± 0.0004 μF 11.3
D ± 0.0030 ± 0.00015 11.4
1 kHz
Cp
±0.00077μF μF
± 0.0001 μF 2.9
D ± 0.0008 ± 0.0001 2.9
0.1 V 1 kHz
Cp
±0.00085μF μF
± 0.0001 μF 3.1
D ± 0.0008 ± 0.0001 3.2