Service manual
34 2
Performance Test
Performance Test
12. Press [Trigger] - Single for a single sweep measurement.
13. Record the multimeter reading and the DC Bias monitor reading, then record them on
the calculation sheet.
14. Calculate the result according to the calculation sheet, then record it on the
performance test record.
15. Perform Step 11 and 14 for all setting in Table 2-5
Table 2-5 DC Bias Monitor Accuracy Test (current) Setting
Bias Level
0 mA
20 mA
100 mA
–20 mA
–100 mA