Service manual

2 21
Performance Test
Performance Test
2. Performance Test
Frequency Accuracy Test
This test checks the frequency accuracy of the internal frequency reference with a
frequency counter and a frequency standard.
Specification
Frequency Accuracy: ± 20 ppm at (23±5) ºC (without Option 1D5)
± 0.13 ppm at 0 ºC to 55 ºC (with Option 1D5)
Test Equipment
Description Recommended Model
Frequency Counter 5334B or 53181A
Frequency Standard 5071A
4TP to BNC Interface Box p/n 04294-61002
BNC(m) -BNC(m) Cable, 61 cm p/n 8120-1839
BNC(m)-BNC(m) Cable, 122 cm p/n 8120-1840
Procedure
1. Connect the test equipment as shown in Figure 2-1. For testing the 4294A with Option
1D5, connect a BNC(m)-BNC(m) cable between the EXT REF Input connector and the
REF OVEN connector on the analyzer rear panel.
NOTE If testing the 4294A without Option 1D5, the frequency standard is not required.