Service manual

234 Appendix C
Error messages
Order of Alphabet
(COPY FILE) key, no file or directory corresponding to the execution of the key was on
the floppy disk.
116 NO STATE/DATA FILES ON MEMORY
Though you attempted to display the list of files and directories on a volatile memory disk
or nonvolatile memory disk (flash memory) on the softkey label by executing the recall
([Recall]) key, file re-save (RE-SAVE FILE) key, file deletion (PURGE FILE) key,
directory change (CHANGE DIRECTORY) key, file copy (COPY FILE) key, and so on,
no file or directory corresponding to the execution of the key was on the volatile memory
disk or nonvolatile memory disk (flash memory).
55 NOT ENOUGH DATA
When data was sent from the controller to the 4294A in the FORM2, FORM3, or FORM5
data transfer format, the amount of the sent binary data was too small, or data items did not
reach the number of points (NOP) set in the 4294A. Set data to be sent or the number of
points of the 4294A correctly.
-120 Numeric data error
Numeric data (including numeric data without a decimal point) causes an error. A numeric
value error other than -121 to -129 occurred.
-128 Numeric data not allowed
A numeric value data element (that did not violate the standard) was received at where the
4294A did not accept any numeric value data elements.
O
140 OSC LEVEL MONITOR NOT VALID
When the oscillator level monitor was not on, you attempted to execute a command to turn
on the marker AC voltage level monitor or the marker AC current level monitor
(MKRMON ACV or MKRMON ACI) and, as a result, the command was ignored. Before
turning on the marker AC voltage level monitor or the marker AC current level monitor,
turn on the AC bias level monitor (command: OMON ON).
P
-220 Parameter error
An error not included in error numbers between -221 and -229 occurred during the analysis
of a program data element. This error occurs, for example, when you attempt to specify an
invalid value for the LOAD correction reference value or the LOAD correction data
(values that are not finite when converted to R-X form impedance values). If this error
occurs, the command is ignored. This error also occurs when you attempt to specify an
invalid LOAD correction reference value from front panel keys.
-108 Parameter not allowed
The number of parameters is larger than that the command requires. For example, though
the “POIN” command requires 1 parameter like “POIN 101,” 2 parameters are added like
POIN 101,1.
89 PERCENT FORMAT NOT SELECTED