Service manual
 214 10 
Post Repair Procedures
POST REPAIR PROCEDURES
POST REPAIR PROCEDURES
Table 10-1 Post Repair Procedures lists the required procedures that must be performed 
after the replacement of an assembly or the EEPROM. These are the recommended 
minimum procedures to ensure that the replacement is successfully completed.
When you replace an assembly or the EEPROM on the A1 CPU, perform the adjustments 
and updating correction constants (CC) in 
Table 10-1. Then perform the operational 
verifications and performance verifications listed in Table 10-1.
For the detailed procedure of the adjustments and updating correction constants, see 
Adjustments Chapter. For the detailed operational verification procedures, see this manual's 
chapter specified in 
Table 10-1. For detailed performance verification procedures, see 
Performance Test Chapter in this manual.
Table 10-1  Post Repair Procedures
Replaced
Assembly or Part
Adjustments
Correction Constants (CC)
Verification
A1 CPU
Firmware Installation
*1
Inspect the Power On Sequence
*2
Internal Test 1: A1 CPU
*1
Internal Test 2: A1 VOLATILE MEMORY
*1
A1 EEPROM Write ID 
All CC Items 
Inspect the Power On Sequence
*2
All Performance Test Items
A2 High Current DC Bias Monitor CC 
DC Bias Level CC 
OSC Level Monitor CC 
OSC Level CC
Inspect the Power On Sequence
*2
External Test 20: ALL ANALOG
*2
OSC Level Accuracy Test 
DC Bias Monitor Accuracy Test 
DC Bias Level Accuracy Test
A3 Hp Amp. Hpot-Amplifier Offset CC 
Relative Impedance CC 
OSC Level Monitor CC 
OSC Level CC 
Impedance Measurement CC
Inspect the Power On Sequence
*2
External Test 20: ALL ANALOG
*2
OSC Level Accuracy Test 
Measurement Accuracy Test
A4 Lc Amp. Null Loop Phase CC 
Relative Impedance CC 
OSC Level Monitor CC 
Impedance Measurement CC
Inspect the Power On Sequence
*2
External Test 20: ALL ANALOG
*2
Measurement Accuracy Test
A5 IV Converter & Lp 
Amp.
Lpot-Mixer Offset CC 
Null Loop Phase CC 
DC Bias Monitor CC 
DC Bias Level CC
Inspect the Power On Sequence
*2
External Test 20:ALL ANALOG
*2
DC Bias Monitor Accuracy Test 
DC Bias Level Accuracy Test 
Measurement Accuracy Test
A6 Vec Gene Null Detector Offset CC 
Null Loop Phase CC
Inspect the Power On Sequence
*2
External Test 20: ALL ANALOG
*2
Measurement Accuracy Test